ScreenWorks Software Overview
72
0112-0109 H
Setup Read Mode
The
Setup Read Mode
table at the top of the
Settings
page displays
settings for up to four read modes that the FLIPR
®
Tetra System can
perform in a single experiment.
Each read mode defines:
•
Reading mode (fluorescence or luminescence)
•
Excitation and emission wavelengths (
i.e.,
a pair of LED banks
and emission filter)
•
Camera gain
•
Camera exposure time
•
Excitation intensity
•
Gate open (ICCD camera only)
Molecular Devices FLIPR
®
Calcium and Membrane Potential Assay Kits
use a single read mode, but ratiometric assays, such as the Voltage
Sensor Probes, require two read modes. In these cases the instrument
alternates between the two read modes for each time point, outputting
two distinct sets of readings, one for each mode. ScreenWorks
Software can then automatically calculate the ratio of the readings from
each mode, for each time point.
If you open a protocol file that uses read modes not available under the
present configuration of your instrument, the unavailable read mode
rows are displayed in red.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...