ScreenWorks Software Overview
92
0112-0109 H
Correction
Use the
Correction
dialog to apply data-correction algorithms, and to
view ratiometric data. These changes affect:
•
Data displayed in the Multi-Well Graph and
Detail Graph
.
•
Data displayed in group statistic table and graphs.
•
Data in files are exported manually or automatically (see
).
•
Data in printed output configured in the
Settings
process page
(see
Plainly, once data are exported, they cannot be changed, so settings
made in this dialog in protocol files are important for these output
options. For viewing data in ScreenWorks Software, however, since raw
data are always kept, correction options selected in protocol files can
be altered when the data are viewed in resulting data files.
The table below provides brief descriptions of the options in this dialog
box; see
Appendix B: Data Processing Algorithms on page 273
description.
Item
Description
Crosstalk
Correction
Only available with Aequorin, ICCD camera. Check this
option to correct all wells for crosstalk from neighboring
wells. The crosstalk function is based on well-to-well
distance and the measured percentage of crosstalk.
Positive Control
Scaling
Check to average each maximum RLU value from the
positive control group and normalize all samples to this
value (set at 100%). This function is useful when graphing a
dose-response curve or when comparing data between
experiments.
Negative
Control
Correction
Check this option to have an average negative control well
value and ratio-to-well value calculated for each sample
interval (time point). This ratio is calculated for each sample
interval and applied to all wells. This function provides a
good correction for signal drift and artifacts.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...