Principles of Operation
12
5000450 D
Site Requirements
Table 1-2
Site Requirements
Temperature
50 to 86ºF (15 to 30ºC)
Humidity
5 to 95% non-condensing
Power
Universal input 50 to 60 Hz, 100 to 240 VAC, 2A max
Space
25.6 inches deep by 16.9 inches wide by 13.4 inches high
(65 cm by 43 cm by 34 cm). An additional 6 inches (15.2
cm) for clearance.
Bench support
Sufficient to support 40 lbs (18 kg) with minimal vibration
Uninterruptible
power supply
(UPS)
We recommend the use of a 325 VA UPS
GenePix_4000B.book Page 12 Friday, October 22, 2010 3:07 PM