ImageXpress Micro High Content Imaging System Options User Guide
40
5014166 D
13.
Click
OK
to close the
Fluidics System Properties
dialog and
return to the
Configure Fluidics Stations
dialog.
14.
If applicable, click
Define Tips
and in the
Define Tips
dialog,
define additional tip types, and then click
Save
.
15.
After you have completed a dispensing cycle, to indicate that you
have reloaded tip racks and plates, click
Reset Tips,
and on the
Reset Tips/Liquid Levels
dialog, click
Reset
for the appropriate
tip trays and plates, and then click
Close
.
Figure 2-8:
Reset Tips/Liquid Levels dialog
Manual Fluidics Control
Do the following procedure to control the fluidics station manually.
CAUTION!
Improper tip definitions can cause the pipette tip to
crash into the sample plate. Molecular Devices recommends
that you do not make changes to the standard tip definitions.
Tip:
The fluidics interface enables you to transfer fluid from one well of
the compound plate to a corresponding well in the sample plate. If you
need to transfer fluid between non-corresponding wells, or if different
actions need to be performed for different wells, record these actions in
a journal (custom routine) to accomplish this. See
. You can add journals as events when you run an
experiment. See
To configure a fluidics experiment on page 43