6.
In the Fluidic Event dialog, select the event type and its
properties. Event types include:
Compound addition
: A Compound addition event, shown in
, draws liquid from the compound
plate and adds it to the sample plate. There is a one-to-one
mapping between the wells on the compound plate and the
sample plate. That is, compound is removed from a well on
the compound plate and applied to the well in the same
position on the sample plate (for example, A1 to A1, A2 to
A2). When you are running mixes that consist of more than a
single cycle,
Mix Dead Volume
specifies an amount of fluid
that is to remain in the tip and not to be returned to the well
during intermediate repeated cycles of draws and dispense.
For more information about compound addition settings,
see the MetaXpress Software application help (press F1
while viewing the dialog).
Figure 2-15:
Fluidic Event dialog Compound addition event
ImageXpressOptionsUserGuide.book Page 46 Wednesday, September 9, 2015 3:54 PM