Mi Microplate Washer User Guide
34
5032022 C
Confirm the Level of the Wash or Rinse Bottle
If the
Liquid Warning
option is enabled in the
Setup
menu, then the software displays a
prompt to make sure that the applicable wash or rinse bottle has sufficient liquid for the
process.
CAUTION!
Running a prime or rinse process without sufficient liquid in the wash or
rinse bottle can damage the instrument. The minimum volume required for a
Prime/Rinse function is 7.5mL+1.5mL (total of 9mL), where 7.5mL is the volume of
liquid in the tube (from Rinse bottle to Dispense head of manifold) and 1.5mL is the
volume dispensed during the Prime/Rinse cycle. We suggest using a little more than
7.5mL + 1.5mL to have enough backup pressure to push liquid through the tube and
dispense the required volume (1.5mL).
To enable the
Liquid Warning
option in the
Setup
menu, see
.
Figure 4-5: The Wash Liquid OK prompt
Button
Action
On the left
If the wash or rinse bottle does not contain enough liquid, press this button to return
to
Prime/Rinse
. See
Priming or Rinsing the Fluidics System on page 31
On the right
If the wash or rinse bottle contains enough liquid, press this button to go to the next
step. If no other options have been enabled, then the manual prime or rinse process
can be started. See
Controlling the Manual Prime or Rinse Process on page 36
The button has no function for this option.
The button has no function for this option.
The button has no function for this option.
Table 4-3: Actions of Control Panel Buttons
Summary of Contents for MultiWash+
Page 1: ...5032022 C August 2019 MultiWash Microplate Washer User Guide ...
Page 20: ...MicroWash Microplate Washer User Guide 20 5032022 C ...
Page 50: ...MicroWash Microplate Washer User Guide 50 5032022 C ...
Page 64: ...MicroWash Microplate Washer User Guide 64 5032022 C ...
Page 144: ...MicroWash Microplate Washer User Guide 144 5032022 C ...