Mi Microplate Washer User Guide
30
5032022 C
Define the Number of Strips to Process
If the
Ask Strip Nr
option is enabled for the selected procedure, then the software displays a
prompt for you to define the number of strips to process. The strips are processed starting
from row A. This can be useful to process a partial microplate or only a few strips in the
procedure.
To enable the
Ask Strip Nr
option in a procedure, see
Defining a Procedure on page 51
.
Figure 3-5: The Number of Strips prompt
Button
Action
On the left
Press this button to return to
Run Procedure
. See
Running a Procedure on page 25
.
On the right
The button has no function for this option.
Press this button to reduce the number of strips to process.
Press this button to increase the number of strips to process.
Press this button to confirm the number of strips to process and then start to run the
selected procedure. See
Monitoring the Procedure on page 30
Table 3-5: Actions of Control Panel Buttons
Monitoring the Procedure
As the selected procedure runs on the instrument, the software displays the name of each
cycle as it is being run.
To stop the procedure, press the
button on the left.