Chapter 9: Regional Picking Processes
5031496 G
111
14. Adjust the Agar Depth, Height, and Maximum Height manually.
In the Agar Depth (mm) field, enter a value in millimeters for the depth that the picking
head descends into the source agar.
The Agar Height (mm) field is read only and displays the height, or thickness, of the
detected agar within the source tray.
The Max. Agar Height (mm) field is read only and displays the maximum agar height for
a long pin head to descend into the source agar
Note: Further refinement for colony detection can be done on a higher-resolution
image in the Feature Selection window.
15. Click Next to go to the next step.
If you are running a routine for white light only, the system captures and processes a
higher-resolution image and then displays the Feature Selection window.
If you are running a routine for white light and fluorescent imaging, the Fluorescent Test
Image window displays.
Adjusting the Fluorescent Test Image
If you are running a routine for white light and fluorescent imaging, the Fluorescent Test Image
window displays after you make adjustments for the white light test image. This option is
available only for instruments with a fluorescence imaging module.
To view the pixel intensity values in the image map to manually set a detection threshold or
detect areas of saturation within the image:
1. Right-click over the image map and select Intensity.
2. Move the cursor over the area of interest in the image map to display the pixel intensity
value.
To adjust the test image:
1. From the Receptacle list, select the number of the receptacle to view and then click the
region to view in the receptacle image below the list. The region is in red.
Summary of Contents for QPix 420
Page 1: ...5031496 G August 2022 QPix 420 Colony Picking System User Guide...
Page 6: ...QPix 420 Colony Picking System User Guide 6 5031496 G...
Page 20: ...QPix 420 Colony Picking System User Guide 20 5031496 G...
Page 30: ...QPix 420 Colony Picking System User Guide 30 5031496 G...
Page 38: ...QPix 420 Colony Picking System User Guide 38 5031496 G...
Page 62: ...QPix 420 Colony Picking System User Guide 62 5031496 G...
Page 138: ...QPix 420 Colony Picking System User Guide 138 5031496 G...
Page 156: ...QPix 420 Colony Picking System User Guide 156 5031496 G...
Page 186: ...QPix 420 Colony Picking System User Guide 186 5031496 G...
Page 194: ...QPix 420 Colony Picking System User Guide 194 5031496 G...
Page 210: ...QPix 420 Colony Picking System User Guide 210 5031496 G...
Page 214: ...QPix 420 Colony Picking System User Guide 214 5031496 G...
Page 218: ...QPix 420 Colony Picking System User Guide 218 5031496 G...
Page 222: ...QPix 420 Colony Picking System User Guide 222 5031496 G...
Page 224: ...QPix 420 Colony Picking System User Guide 224 5031496 G...