QPix 420 Colony Picking System User Guide
136
5031496 G
Viewing the Regional Picking Progress
While the regional picking routine is running, the Regional Picking Progress window displays a
summary of the routine.
Start Time shows the time that the picking of the colonies began.
Source Barcode displays the barcode or identifier of the source receptacle.
Pin is the picking pin being used for the current picking operation.
Copy Plate No is the barcode of the copy receptacle that is being inoculated, if you select
Make Copy for the routine.
Destination Barcode is the barcode of the destination receptacle that is being inoculated.
Destination Offset is the reference well for where the pins are lowered. This value changes
for 384-well plates.
Colonies Picked is the number of colonies picked so far.
Colonies to Pick is the total number of colonies to be picked.
Calculate Remaining Time indicates remaining time in the routine.
Current Action displays what the system is doing, such as Pick when the system is picking.
To turn the light table on or off, click Light Table On or Light Table Off.
To safely pause the routine, click Pause.
After all the colonies are picked from the source receptacles and delivered to the destination
plates, the Finished Picking Current Batch dialog displays.
Continuing or Ending the Regional Picking Routine
After all the colonies are picked from the source receptacles and delivered to the destination
plates, the Finished Picking Current Batch dialog displays.
If there are no more source receptacles to pick, click Finish Picking and then click OK.
To continue picking more source receptacles:
1. Select whether or not to review images of the new receptacles.
To review images and select colonies for the new source receptacles, click Continue
With Image Review.
To use the current image settings for the next batch of source receptacles, click
Continue Without Image Review.
2. From the Select Number of Positions to Use, select the number of receptacle holders to
load on the instrument deck.
3. Click OK.
4. When the Please Load Source window displays, replace the picked receptacles with the
new receptacles on the instrument deck.
5. Close the instrument door.
6. Click Next to continue running the picking routine.
If you click Continue With Image Review, the system takes a white light test image of
the source receptacles and then displays the Test Image window. See
Blue-White Regional White Light Test Image on page 129
If you click Continue Without Image Review, the system captures and processes a
higher-resolution image and then displays the Feature Selection window. See
Summary of Contents for QPix 420
Page 1: ...5031496 G August 2022 QPix 420 Colony Picking System User Guide...
Page 6: ...QPix 420 Colony Picking System User Guide 6 5031496 G...
Page 20: ...QPix 420 Colony Picking System User Guide 20 5031496 G...
Page 30: ...QPix 420 Colony Picking System User Guide 30 5031496 G...
Page 38: ...QPix 420 Colony Picking System User Guide 38 5031496 G...
Page 62: ...QPix 420 Colony Picking System User Guide 62 5031496 G...
Page 138: ...QPix 420 Colony Picking System User Guide 138 5031496 G...
Page 156: ...QPix 420 Colony Picking System User Guide 156 5031496 G...
Page 186: ...QPix 420 Colony Picking System User Guide 186 5031496 G...
Page 194: ...QPix 420 Colony Picking System User Guide 194 5031496 G...
Page 210: ...QPix 420 Colony Picking System User Guide 210 5031496 G...
Page 214: ...QPix 420 Colony Picking System User Guide 214 5031496 G...
Page 218: ...QPix 420 Colony Picking System User Guide 218 5031496 G...
Page 222: ...QPix 420 Colony Picking System User Guide 222 5031496 G...
Page 224: ...QPix 420 Colony Picking System User Guide 224 5031496 G...