Selecting the Head and Sanitizing Options
To select the head and sanitizing options:
1. In the Head and Sanitise window, from the Picking Head list, select the head to use for the
picking routine.
2. In the Pin Columns (Rows) Before Inoculation list, select the number of pins to use to pick in
each column or row before transferring the picked colonies to the destination plate.
For the fastest transfer rate, select All to use all available picking pins to pick colonies
before inoculating the destination plate.
For smaller picking groups, the available selections depend on the Deposit Order selection
made in the Destination Options window.
If you select By Columns, you can pick colonies using from 1 to 11 columns.
If you select By Rows, you can pick colonies using from 1 to 7 rows.
To run the wash routine from the Sanitise profile after each partial pick and inoculation,
select the Wash Between Partial Inoculation check box. Clearing this check box waits until
all pins have been used for picking and inoculation before running the wash routine.
3. In the Destination field, enter the distance in millimeters above the bottom of the
destination plate wells to dip the pins for the inoculation.
If you select Make Copy for the destination plate, enter a value for the copy.
4. From the Sanitise Profile list, select the Sanitise profile to use for the picking routine.
If the available profiles are not suitable for the picking routine, exit the picking process and
create a new Sanitise profile. See
Creating and Editing Sanitise Profiles on page 33
.
5. Click Next to define options for the source receptacle.
Setting Source Receptacle Options
To set source receptacle options:
1. In the Source window, from the Holder list, select the type of holder used to hold the source
receptacles on the instrument deck.
2. From the Receptacle list, select the type of receptacle used for the source. Only the
receptacles applicable for the holder display in this list.
The preview image displays a representation of the holder and receptacle.
3. In the Positions field, enter the number of receptacles to use for picking.
4. In the Offset field, enter the position of the first receptacle to use for picking.
The preview image displays the defined locations of the receptacles.
Note: If you are using Validation barcodes, only one source receptacle shows, and
the Positions and Offset fields are not available.
5. In the Picking Depth Into Agar (mm) field, enter a value in millimeters to define the depth
that the pins are to be inserted in the agar for picking.
6. Select the Limit Max. Number of Features Per Position check box and then in the Max.
Number of Features Per Position field, enter the maximum number of colonies that can be
picked from a single QTray or Petri dish.
Note: Setting a limit for the maximum number of colonies in the Source window
prevents selecting a maximum number in the Feature Selection window when
running the picking routine. See
Selecting Colonies for Picking on page 53
.
7. Click Next to view a summary of the settings.
Summary of Contents for QPix 420
Page 1: ...5031496 G August 2022 QPix 420 Colony Picking System User Guide...
Page 6: ...QPix 420 Colony Picking System User Guide 6 5031496 G...
Page 20: ...QPix 420 Colony Picking System User Guide 20 5031496 G...
Page 30: ...QPix 420 Colony Picking System User Guide 30 5031496 G...
Page 38: ...QPix 420 Colony Picking System User Guide 38 5031496 G...
Page 62: ...QPix 420 Colony Picking System User Guide 62 5031496 G...
Page 138: ...QPix 420 Colony Picking System User Guide 138 5031496 G...
Page 156: ...QPix 420 Colony Picking System User Guide 156 5031496 G...
Page 186: ...QPix 420 Colony Picking System User Guide 186 5031496 G...
Page 194: ...QPix 420 Colony Picking System User Guide 194 5031496 G...
Page 210: ...QPix 420 Colony Picking System User Guide 210 5031496 G...
Page 214: ...QPix 420 Colony Picking System User Guide 214 5031496 G...
Page 218: ...QPix 420 Colony Picking System User Guide 218 5031496 G...
Page 222: ...QPix 420 Colony Picking System User Guide 222 5031496 G...
Page 224: ...QPix 420 Colony Picking System User Guide 224 5031496 G...