Creating and Editing Regional Picking Processes
The procedures to create and edit standard or regional picking processes are similar. The
regional picking process has more options available for defining the regions for picking.
If this is the first regional picking process, you must edit or create a Sanitise profile to use with
the regional picking process. See
Creating and Editing Sanitise Profiles on page 33
.
Creating and editing a regional picking process involves the following procedures:
Opening the Regional Picking Window on page 100
Selecting Regional Picking Routines on page 100
Selecting Barcode Options and Filter Pairs for Fluorescent Imaging on page 102
Setting Destination Plate Options on page 104
Selecting the Regional Picking Source on page 104
Selecting the Head and Sanitizing Options on page 106
Viewing the Settings Summary on page 106
Changing the Picking Head on page 107
Opening the Regional Picking Window
To open the Regional Picking window:
1. From the Navigation window under Picking Processes, double-click the Regional Picking
icon.
2. In the Regional Picking window, click Picking Type to view the picking type options.
Select White Light to use only white light to illuminate and identify the colonies to pick.
Select White Light And Fluorescent to use white light to illuminate and identify
colonies, and fluorescent light to determine the colonies to pick. This option is available
only for instruments with a fluorescence imaging module.
3. Click Apply.
4. Click Start to home the drives and display the Routines window.
Selecting Regional Picking Routines
To select a Regional Picking routine:
1. In the Routines window, select to create a new routine or to edit a routine.
To create a new routine, click New Routine.
To edit a routine, click Run Existing Routine and then select the name of the routine
from the Select Routine list.
To run the routine without making changes, select the Skip Steps check box.
2. Click Next to go to the next step.
If you select to create a new routine or to edit a routine, see
and Filter Pairs for Fluorescent Imaging on page 102
If you select the Skip Steps check box to run the routine without making changes, see
Viewing the Settings Summary on page 106
Note: You can import, export, and delete routines in the Routines window.
Summary of Contents for QPix 420
Page 1: ...5031496 G August 2022 QPix 420 Colony Picking System User Guide...
Page 6: ...QPix 420 Colony Picking System User Guide 6 5031496 G...
Page 20: ...QPix 420 Colony Picking System User Guide 20 5031496 G...
Page 30: ...QPix 420 Colony Picking System User Guide 30 5031496 G...
Page 38: ...QPix 420 Colony Picking System User Guide 38 5031496 G...
Page 62: ...QPix 420 Colony Picking System User Guide 62 5031496 G...
Page 138: ...QPix 420 Colony Picking System User Guide 138 5031496 G...
Page 156: ...QPix 420 Colony Picking System User Guide 156 5031496 G...
Page 186: ...QPix 420 Colony Picking System User Guide 186 5031496 G...
Page 194: ...QPix 420 Colony Picking System User Guide 194 5031496 G...
Page 210: ...QPix 420 Colony Picking System User Guide 210 5031496 G...
Page 214: ...QPix 420 Colony Picking System User Guide 214 5031496 G...
Page 218: ...QPix 420 Colony Picking System User Guide 218 5031496 G...
Page 222: ...QPix 420 Colony Picking System User Guide 222 5031496 G...
Page 224: ...QPix 420 Colony Picking System User Guide 224 5031496 G...