QPix 420 Colony Picking System User Guide
106
5031496 G
Selecting the Head and Sanitizing Options
To select the head and sanitizing options:
1. In the Head and Sanitise window, from the Picking Head list, select the head to use for the
picking routine.
2. From the Pin Columns (Rows) Before Inoculation list, select the number of pins to use for
picking in each column or row before transferring the picked colonies to the destination
plate.
For the fastest transfer rate, select All to use all available picking pins to pick colonies
before inoculating the destination plate.
For smaller picking groups, the available selections depend on the Deposit Order selection
made in the Destination/Source Options window.
If you select By Columns, you can pick colonies using from 1 to 11 columns.
If you select By Rows, you can pick colonies using from 1 to 7 rows.
To run the wash routine from the Sanitise profile after each partial pick and inoculation,
select the Wash Between Partial Inoculation check box. Clear this check box to wait until all
pins are used for picking and inoculation before running the wash routine.
3. In the Destination field, enter the distance in millimeters above the bottom of the
destination plate wells to dip the pins for the inoculation.
If you select Make Copy for the destination plate, enter a value for the copy.
4. From the Sanitise Profile list, select the Sanitise profile to use for the picking routine.
If the available profiles are not suitable for the regional picking routine, exit the regional
picking process and create a new Sanitise profile. See
Creating and Editing Sanitise Profiles
5. Click Next to view a summary of the settings.
Viewing the Settings Summary
The Settings Summary window contains a full summary of the picking routine settings.
Review the summary details to make sure that the settings and options are configured correctly
for the regional picking routine. If you must make changes, click Back until you return to the
window where the changes can be made.
To print the summary, click Print.
To run the regional picking routine, click Next.
Summary of Contents for QPix 420
Page 1: ...5031496 G August 2022 QPix 420 Colony Picking System User Guide...
Page 6: ...QPix 420 Colony Picking System User Guide 6 5031496 G...
Page 20: ...QPix 420 Colony Picking System User Guide 20 5031496 G...
Page 30: ...QPix 420 Colony Picking System User Guide 30 5031496 G...
Page 38: ...QPix 420 Colony Picking System User Guide 38 5031496 G...
Page 62: ...QPix 420 Colony Picking System User Guide 62 5031496 G...
Page 138: ...QPix 420 Colony Picking System User Guide 138 5031496 G...
Page 156: ...QPix 420 Colony Picking System User Guide 156 5031496 G...
Page 186: ...QPix 420 Colony Picking System User Guide 186 5031496 G...
Page 194: ...QPix 420 Colony Picking System User Guide 194 5031496 G...
Page 210: ...QPix 420 Colony Picking System User Guide 210 5031496 G...
Page 214: ...QPix 420 Colony Picking System User Guide 214 5031496 G...
Page 218: ...QPix 420 Colony Picking System User Guide 218 5031496 G...
Page 222: ...QPix 420 Colony Picking System User Guide 222 5031496 G...
Page 224: ...QPix 420 Colony Picking System User Guide 224 5031496 G...