5031496 G
153
Chapter 12: Manage Regional Tray Processes
The QPix 420 Colony Picking System can pick colonies from receptacles using either standard
picking or regional picking. You can set up control wells for the destination plates before
running a picking process.
The following options are available in the Navigation window under Picking Processes:
Picking picks colonies from receptacles using the standard process. See
Blue-White Picking picks blue or white colonies from receptacles using the standard
process. See
Blue-White Picking Processes on page 63
Zone of Inhibition Detection Picking picks colonies with detectable zones of inhibition from
receptacles using the standard process. See
Zone of Inhibition Detection Picking
Regional Picking picks colonies from receptacles using the regional process. See
.
Blue-White Regional Picking picks blue or white colonies from receptacles using the
regional process.
Blue-White Regional Picking Processes on page 119
Control Plate Creation creates a batch of plates that contain control samples. These can be
created by picking colonies from source receptacles into destination wells. See
Plate Creation Processes on page 139
Manage Regional Trays adds, modifies, or removes new regional tray source definitions to
the regional picking source type database, as described in this chapter, which can then be
selected for use in the regional picking process.
For the instruments licensed for Regional Picking, you can also define new custom plate
receptacles to be picked from using an OmniTray holder in a Regional Picking process.
Create New Regional Trays
Your instrument must be licensed for Regional Picking to use this feature.
The new plate source tray must be able to be loaded within an OmniTray holder that fits on the
light table of the instrument. The OmniTray supports trays that conform to the external
dimensions within the ANSI/SLAS specification for plates with the following standard
dimensions:
Length: 127.76 mm ± 0.25 mm (5.0299 inches ± 0.0098 inches)
Width: 85.48 mm ± 0.25 mm (3.3654 inches ± 0.0098 inches)
12
Summary of Contents for QPix 420
Page 1: ...5031496 G August 2022 QPix 420 Colony Picking System User Guide...
Page 6: ...QPix 420 Colony Picking System User Guide 6 5031496 G...
Page 20: ...QPix 420 Colony Picking System User Guide 20 5031496 G...
Page 30: ...QPix 420 Colony Picking System User Guide 30 5031496 G...
Page 38: ...QPix 420 Colony Picking System User Guide 38 5031496 G...
Page 62: ...QPix 420 Colony Picking System User Guide 62 5031496 G...
Page 138: ...QPix 420 Colony Picking System User Guide 138 5031496 G...
Page 156: ...QPix 420 Colony Picking System User Guide 156 5031496 G...
Page 186: ...QPix 420 Colony Picking System User Guide 186 5031496 G...
Page 194: ...QPix 420 Colony Picking System User Guide 194 5031496 G...
Page 210: ...QPix 420 Colony Picking System User Guide 210 5031496 G...
Page 214: ...QPix 420 Colony Picking System User Guide 214 5031496 G...
Page 218: ...QPix 420 Colony Picking System User Guide 218 5031496 G...
Page 222: ...QPix 420 Colony Picking System User Guide 222 5031496 G...
Page 224: ...QPix 420 Colony Picking System User Guide 224 5031496 G...