QPix 420 Colony Picking System User Guide
202
5031496 G
Aligning the Camera
To ensure accurate picking, the camera must be calibrated and correctly aligned to get pin-to-
spot precision, relating the image pixel coordinates with the instrument x and y coordinates. Do
this process whenever the head is returned to the actuator or if the actuator is hit during a
maintenance operation, as this can have a negative effect on picking precision.
For the camera alignment procedure, make sure that you use a standard 96-pin picking head.
To align the camera:
1. From the Navigation window under Utility Processes, double-click the Camera Alignment
Process icon, and wait for the Camera Alignment Process window to display.
2. If you do not have a standard 96-pin picking head installed, click Move to Load Position.
See
.
After you install the standard 96-pin picking head, click Move to Park Position and then
click Next.
3. Select the type of source plate that you must place on the light table from the Holder and
Receptacle fields.
4. Change the Offset to specify an area of the source plate to use for the camera alignment
process other than the default position.
5. Click Next.
6. Place the source plate with agar on the light table as instructed and then click Next.
7. Make sure that the lid is removed and that the source plate is in the position indicated and
then click Next.
Summary of Contents for QPix 420
Page 1: ...5031496 G August 2022 QPix 420 Colony Picking System User Guide...
Page 6: ...QPix 420 Colony Picking System User Guide 6 5031496 G...
Page 20: ...QPix 420 Colony Picking System User Guide 20 5031496 G...
Page 30: ...QPix 420 Colony Picking System User Guide 30 5031496 G...
Page 38: ...QPix 420 Colony Picking System User Guide 38 5031496 G...
Page 62: ...QPix 420 Colony Picking System User Guide 62 5031496 G...
Page 138: ...QPix 420 Colony Picking System User Guide 138 5031496 G...
Page 156: ...QPix 420 Colony Picking System User Guide 156 5031496 G...
Page 186: ...QPix 420 Colony Picking System User Guide 186 5031496 G...
Page 194: ...QPix 420 Colony Picking System User Guide 194 5031496 G...
Page 210: ...QPix 420 Colony Picking System User Guide 210 5031496 G...
Page 214: ...QPix 420 Colony Picking System User Guide 214 5031496 G...
Page 218: ...QPix 420 Colony Picking System User Guide 218 5031496 G...
Page 222: ...QPix 420 Colony Picking System User Guide 222 5031496 G...
Page 224: ...QPix 420 Colony Picking System User Guide 224 5031496 G...