Chapter 17: Maintenance
5031496 G
205
To replace a fuse:
WARNING! HIGH VOLTAGE
Power off the instrument and disconnect the power
cord before you do maintenance procedures that require removal of a panel or cover
or disassembly of an interior instrument component.
1. Switch the power switch on the front of the instrument to the off position.
2. Unplug the power cord from the power port.
3. Use a small slot-head screwdriver to turn the fuse carrier counter-clockwise and then pull
the fuse carrier out to expose the fuse.
4. Use the screwdriver to gently lift the old fuse from the carrier.
5. Gently place a new fuse into the carrier by hand.
6. Slide the fuse carrier back into the instrument.
7. Use the screwdriver to turn the carrier clockwise until it is snug, but do not over tighten it.
8. Plug the power cord into the power port.
9. Turn on the power to the instrument.
Note: If the instrument still does not power on after changing the fuse, contact technical
support.
Draining the Compressor
The following procedure is for the optional DynAir DA7001CS air compressor. The compressor
collects water in the internal air tank. This water must be emptied into the externally mounted
fluid collection bottle. Draining frequency depends on instrument use:
High use - drain once a day
Low use - drain once a week
To drain the air tank:
1. Verify that the compressor pressure is 105 psi.
2. Turn off the air compressor.
Summary of Contents for QPix 420
Page 1: ...5031496 G August 2022 QPix 420 Colony Picking System User Guide...
Page 6: ...QPix 420 Colony Picking System User Guide 6 5031496 G...
Page 20: ...QPix 420 Colony Picking System User Guide 20 5031496 G...
Page 30: ...QPix 420 Colony Picking System User Guide 30 5031496 G...
Page 38: ...QPix 420 Colony Picking System User Guide 38 5031496 G...
Page 62: ...QPix 420 Colony Picking System User Guide 62 5031496 G...
Page 138: ...QPix 420 Colony Picking System User Guide 138 5031496 G...
Page 156: ...QPix 420 Colony Picking System User Guide 156 5031496 G...
Page 186: ...QPix 420 Colony Picking System User Guide 186 5031496 G...
Page 194: ...QPix 420 Colony Picking System User Guide 194 5031496 G...
Page 210: ...QPix 420 Colony Picking System User Guide 210 5031496 G...
Page 214: ...QPix 420 Colony Picking System User Guide 214 5031496 G...
Page 218: ...QPix 420 Colony Picking System User Guide 218 5031496 G...
Page 222: ...QPix 420 Colony Picking System User Guide 222 5031496 G...
Page 224: ...QPix 420 Colony Picking System User Guide 224 5031496 G...