Chapter 4: Starting and Setting Up the Instrument
5031496 G
27
Preparing to Run Processes
Molecular Devices recommends regular and thorough cleaning and maintenance of the
QPix 420 system Instrument to ensure that it continues to function correctly. See
Before running a process, do the following procedures.
1. Manually clean the instrument interior. See
Cleaning the Instrument on page 196
2. Install and fill the wash baths with a suitable quantity of the defined solutions.
Note: Before running a process that uses the wash baths, make sure that you verify
the locations and solutions of the wash baths and also the wash cycles that are
required for their use. Always top off your wash baths with their defined cleaning
fluids and then wipe down the surfaces of the instrument interior.
3. Start the instrument and the software.
4. Install the correct picking head for your application. See
.
5. Do a Sanitise process. See
Running Sanitise Processes on page 31
6. Do a UV Sanitise process. See
Running UV Sanitise Processes on page 32
7. Prepare plates, QTrays, OmniTrays, or Petri dishes based on the requirements for your
application.
8. Select and run the desired process. For information about selecting, defining, and running
processes, see the following topics:
Picking Processes on page 39
Regional Picking Processes on page 99
Control Plate Creation Processes on page 139
Replication Processes on page 157
Rearraying Processes on page 167
Gridding Processes on page 175
Setting Up Wash Baths
The Instrument contains three wash baths on the right side of the instrument bed. The wash
baths are used for the Sanitise process and the Sanitise profiles used with most other
processes. See
Sanitizing the Instrument Interior on page 31
.
Note: Before running a process that uses the wash baths, make sure that you verify the
locations and solutions of the wash baths and also the wash cycles that are required for
their use. Always top off your wash baths with their defined cleaning fluids and then wipe
down the surfaces of the instrument interior.
Fill the wash baths with fluid until the brushes are covered and configure the baths correctly.
For a typical picking process, put the following solutions in the indicated wash baths:
Wash Bath 1 (furthest from front): 70% ethanol
Wash Bath 2 (middle wash bath): Sterile water
Summary of Contents for QPix 420
Page 1: ...5031496 G August 2022 QPix 420 Colony Picking System User Guide...
Page 6: ...QPix 420 Colony Picking System User Guide 6 5031496 G...
Page 20: ...QPix 420 Colony Picking System User Guide 20 5031496 G...
Page 30: ...QPix 420 Colony Picking System User Guide 30 5031496 G...
Page 38: ...QPix 420 Colony Picking System User Guide 38 5031496 G...
Page 62: ...QPix 420 Colony Picking System User Guide 62 5031496 G...
Page 138: ...QPix 420 Colony Picking System User Guide 138 5031496 G...
Page 156: ...QPix 420 Colony Picking System User Guide 156 5031496 G...
Page 186: ...QPix 420 Colony Picking System User Guide 186 5031496 G...
Page 194: ...QPix 420 Colony Picking System User Guide 194 5031496 G...
Page 210: ...QPix 420 Colony Picking System User Guide 210 5031496 G...
Page 214: ...QPix 420 Colony Picking System User Guide 214 5031496 G...
Page 218: ...QPix 420 Colony Picking System User Guide 218 5031496 G...
Page 222: ...QPix 420 Colony Picking System User Guide 222 5031496 G...
Page 224: ...QPix 420 Colony Picking System User Guide 224 5031496 G...