Chapter 9: Regional Picking Processes
5031498 G
125
Selecting the Regional Picking Source
Note: The types of source receptacles used for regional picking are limited. Depending
on the configuration of your system, you might have only one option in the Holder list, in
which case the Receptacle option is preset and cannot be changed. If the type of source
receptacle that you need is not listed, contact Molecular Devices to add a new source
receptacle type to the list. See
.
To select the regional picking source:
1. In the Source window, from the Holder list, select the type of holder used to hold the source
receptacles on the instrument deck.
2. From the Receptacle list, select the type of receptacle used for the source. Only the
receptacles applicable for the holder display in this list.
The preview image displays a representation of the holder and receptacle.
3. In the Positions field, enter the number of receptacles to use for picking.
4. In the Offset field, enter the position of the first receptacle to use for picking.
The preview image displays the defined locations of the receptacles.
Note: If you are using Validation barcodes, only one source receptacle displays, and
the Positions and Offset fields are not available.
5. In the Picking Depth Into Agar (mm) field, enter a value in millimeters to define the depth
that the pins are to be inserted in the agar for picking.
6. Click Next to the source and destination options.
Setting Regional Picking Source and Destination Options
To set the regional picking source and destination options:
1. In the Destination/Source Options window, select the Limit Max. Number of Features Per
Region check box and then in the Number of Features Per Region field, enter the
maximum number of colonies that can be picked from a single source receptacle.
Note: Setting a limit for the maximum number of colonies in the Destination/Source
Options window prevents selecting a maximum number in the Feature Selection
window. See
Select the Reserve Wells for Regions check box to leave wells empty if the number of
colonies per region is not reached. For example, if you select to pick 8 colonies per
region, but only 6 colonies were eligible to pick, wells 7 and 8 would be left blank.
Select the Skip Region If Number of Features Not Found check box to ignore the
region if the target number of features in Number of Features Per Region is not met.
Select the Keep Picked Regions Together check box to have colonies that are picked
from the same region delivered to the same destination plate.
With the Reserve Wells for Region check box selected, select the Skip Region If
Picking Count is Zero check box to skip regions that have a picking count of zero.
2. Select the Deposit Order to deposit the picked colonies in the destination plates By
Columns or By Rows.
Summary of Contents for QPix 450
Page 1: ...5031498 G August 2022 QPix 450 or QPix 460 Colony Picking System User Guide ...
Page 6: ...QPix 450 or QPix 460 Colony Picking System User Guide 6 5031498 G ...
Page 22: ...QPix 450 or QPix 460 Colony Picking System User Guide 22 5031498 G ...
Page 56: ...QPix 450 or QPix 460 Colony Picking System User Guide 56 5031498 G ...
Page 102: ...QPix 450 or QPix 460 Colony Picking System User Guide 102 5031498 G ...
Page 158: ...QPix 450 or QPix 460 Colony Picking System User Guide 158 5031498 G ...
Page 176: ...QPix 450 or QPix 460 Colony Picking System User Guide 176 5031498 G ...
Page 218: ...QPix 450 or QPix 460 Colony Picking System User Guide 218 5031498 G ...
Page 226: ...QPix 450 or QPix 460 Colony Picking System User Guide 226 5031498 G ...
Page 250: ...QPix 450 or QPix 460 Colony Picking System User Guide 250 5031498 G ...
Page 256: ...QPix 450 or QPix 460 Colony Picking System User Guide 256 5031498 G ...