Chapter 9: Regional Picking Processes
5031498 G
129
Running Regional Picking Processes
After a regional picking routine has been configured, you can run the process on the
instrument. If you have not configured the regional picking routine to run, you must create a
new regional picking routine or edit a routine. See
Creating and Editing Regional Picking
.
Note: Before you run a regional picking process, it is important that you do the cleaning
and set up procedures in
Preparing to Run Processes on page 29
Some steps that are included in these procedures might not be available, depending on the
features included with the instrument and license.
To run a configured regional picking routine:
1. Open the Regional Picking window. See
Opening the Regional Picking Window on page
.
2. Select the picking routine to run. See
Selecting Regional Picking Routines on page 120
If you do not need to make changes to the routine, select the Skip Steps check box before
clicking Next.
3. Review the configured settings for the routine. See
Viewing the Settings Summary on page
4. In the Settings Summary window, click Next.
5. When the Please Load Source window displays, load the source receptacles in the correct
locations on the instrument deck.
6. Close the instrument door.
7. Click Next to take a white light test image of the source receptacles.
Adjusting the White Light Test Image
In the Test Image window, view detected features as potential colonies to pick for inoculation of
the destination plates. To make sure that only colonies are detected, adjust the test image
using the Exposure and Gain settings, the Detection settings, the Display settings, the Debris
settings, and the Depth settings. Each detected feature displays within a yellow ring.
To set a detection threshold or detect areas of saturation within the image:
1. Right-click over the image map and select Intensity.
2. Move the cursor over the area of interest in the image map to display the pixel intensity
value.
Summary of Contents for QPix 450
Page 1: ...5031498 G August 2022 QPix 450 or QPix 460 Colony Picking System User Guide ...
Page 6: ...QPix 450 or QPix 460 Colony Picking System User Guide 6 5031498 G ...
Page 22: ...QPix 450 or QPix 460 Colony Picking System User Guide 22 5031498 G ...
Page 56: ...QPix 450 or QPix 460 Colony Picking System User Guide 56 5031498 G ...
Page 102: ...QPix 450 or QPix 460 Colony Picking System User Guide 102 5031498 G ...
Page 158: ...QPix 450 or QPix 460 Colony Picking System User Guide 158 5031498 G ...
Page 176: ...QPix 450 or QPix 460 Colony Picking System User Guide 176 5031498 G ...
Page 218: ...QPix 450 or QPix 460 Colony Picking System User Guide 218 5031498 G ...
Page 226: ...QPix 450 or QPix 460 Colony Picking System User Guide 226 5031498 G ...
Page 250: ...QPix 450 or QPix 460 Colony Picking System User Guide 250 5031498 G ...
Page 256: ...QPix 450 or QPix 460 Colony Picking System User Guide 256 5031498 G ...