5031498 G
Calculate Remaining Time indicates remaining time in the routine.
Current Action displays what the system is currently doing, such as Pick when the system is
picking.
To turn the light table on or off, click Light Table On or Light Table Off.
To safely pause the routine, click Pause.
After all the colonies are picked from the source receptacles and delivered to the destination
plates, the Finished Picking Current Batch dialog displays.
Continuing or Ending the Regional Picking Routine
After all the colonies are picked from the source receptacles and delivered to the destination
plates, the Finished Picking Current Batch dialog displays.
If there are no more source receptacles to be picked, click Finish Picking and then click OK.
To continue picking more source receptacles:
1. Select whether or not to review images of the new receptacles.
To review images and select colonies for the new source receptacles, click Continue
With Image Review.
To use the current image settings for the next batch of source receptacles, click
Continue Without Image Review.
2. From the Select Number of Positions to Use, select the number of receptacle holders to
load on the instrument deck.
3. Click OK.
4. When the Please Load Source window displays, replace the picked receptacles with the
new receptacles on the instrument deck.
5. Close the instrument door.
6. Click Next to continue running the picking routine.
If you click Continue With Image Review, the system takes a white light test image of
the source receptacles and then displays the Test Image window. See
White Light Test Image on page 129
.
If you click Continue Without Image Review, the system captures and processes a
higher-resolution image and then displays the Feature Selection window. See
.
Viewing the Regional Picking Summary
After all regional picking routines complete, the Regional Picking Summary window displays,
showing the number of source colonies picked, the number of destination plates used, and
missing source receptacles.
To save this information in .csv format, click Export.
To view details of all activities related to the source and destination receptacles, click Details.
To save the detailed information in .csv format, click Export.
To close the picking details and return to the Regional Picking Summary window, click
Close.
In the Regional Picking Summary window, click Next and then click Finish to return to the
Regional Picking window.
Summary of Contents for QPix 450
Page 1: ...5031498 G August 2022 QPix 450 or QPix 460 Colony Picking System User Guide ...
Page 6: ...QPix 450 or QPix 460 Colony Picking System User Guide 6 5031498 G ...
Page 22: ...QPix 450 or QPix 460 Colony Picking System User Guide 22 5031498 G ...
Page 56: ...QPix 450 or QPix 460 Colony Picking System User Guide 56 5031498 G ...
Page 102: ...QPix 450 or QPix 460 Colony Picking System User Guide 102 5031498 G ...
Page 158: ...QPix 450 or QPix 460 Colony Picking System User Guide 158 5031498 G ...
Page 176: ...QPix 450 or QPix 460 Colony Picking System User Guide 176 5031498 G ...
Page 218: ...QPix 450 or QPix 460 Colony Picking System User Guide 218 5031498 G ...
Page 226: ...QPix 450 or QPix 460 Colony Picking System User Guide 226 5031498 G ...
Page 250: ...QPix 450 or QPix 460 Colony Picking System User Guide 250 5031498 G ...
Page 256: ...QPix 450 or QPix 460 Colony Picking System User Guide 256 5031498 G ...