5031498 G
177
Chapter 13: Plating Processes
The QPix 460 Colony Picking System uses the plating process to aspirate liquid samples from
source plates and then to dispense the liquid on the surface of agar in a 48-region divided
QTray. The plating head spreads the sample from the source plate in defined patterns across
the agar within each region of the destination receptacle. The plating process prepares QTrays
for regional picking of the formed colonies. See
Regional Picking Processes on page 119
Before you create a plating routine, make sure the pattern you want is available in the Plating
Pattern Editor. Each pattern contains a minimum of one path, and multiple paths can be created
to spread the dispensed liquid in the region. See
Creating and Editing Plating Patterns on page
.
Note: Before you run a plating process, it is important that you do the cleaning and set
up procedures in
Preparing to Run Processes on page 29
If this is the first plating process, you must edit or create a Sanitise profile to use with the plating
process. See
Creating and Editing Sanitise Profiles on page 51
If you are unsure whether the liquid handling unit is aspirating and dispensing correctly, do a
calibration test. See
Calibrating Aspirated and Dispensed Liquid Volumes on page 186
Note: Plating processes are available for the QPix 460 System only.
Creating and Editing Plating Processes
If this is the first plating process, you must edit or create a Sanitise profile to use with the plating
process. See
Creating and Editing Sanitise Profiles on page 51
Creating and editing a plating process involves the following procedures:
Opening the Plating Window on page 177
Selecting Plating Routines on page 178
Selecting Barcode Options on page 179
Setting Source and Destination Options on page 181
Setting Volume and Depth Options on page 181
Selecting the Sanitise Options on page 182
Selecting the Source Stackers on page 182
Viewing the Settings Summary on page 182
After you create the routine, you can close the process and run it later, or you can run it
immediately.
Note: Some steps that are included in these procedures might not be available,
depending on the features included with the instrument and license.
Opening the Plating Window
To open the Plating window:
1. From the Navigation window under Plating Processes, double-click the Plating icon.
2. In the Plating window, click Start to home the drives and display the Routines window.
13
Summary of Contents for QPix 450
Page 1: ...5031498 G August 2022 QPix 450 or QPix 460 Colony Picking System User Guide ...
Page 6: ...QPix 450 or QPix 460 Colony Picking System User Guide 6 5031498 G ...
Page 22: ...QPix 450 or QPix 460 Colony Picking System User Guide 22 5031498 G ...
Page 56: ...QPix 450 or QPix 460 Colony Picking System User Guide 56 5031498 G ...
Page 102: ...QPix 450 or QPix 460 Colony Picking System User Guide 102 5031498 G ...
Page 158: ...QPix 450 or QPix 460 Colony Picking System User Guide 158 5031498 G ...
Page 176: ...QPix 450 or QPix 460 Colony Picking System User Guide 176 5031498 G ...
Page 218: ...QPix 450 or QPix 460 Colony Picking System User Guide 218 5031498 G ...
Page 226: ...QPix 450 or QPix 460 Colony Picking System User Guide 226 5031498 G ...
Page 250: ...QPix 450 or QPix 460 Colony Picking System User Guide 250 5031498 G ...
Page 256: ...QPix 450 or QPix 460 Colony Picking System User Guide 256 5031498 G ...