QPix 450 or QPix 460 Colony Picking System User Guide
44
5031498 G
Setting Up Stacker Lanes for Static Holder Usage
You can bypass using the high-throughput automatic stacker lanes for the Picking, Rearraying,
Replicating, Gridding, and Plating (QPix 460only) processes by selecting the Use Static Holder
check box during the Stackers, Destination Stackers, or Head/Stackers step of your routine set
up. You are prompted to manually load the plate at the end of the stacker lane in the routine set
up.
The static holder functionality is available through the purchase of the optional
Static Plate
Holder Software License for QPix® 450
/460 Colony Picking Systems that is compatible with
QPix Software Version 2.2 and newer.
The Use Static Holder option is used with non-configured standard plates in a standard plate
stacker lane. For a deep well plate stacker lane, the Use Static Holder option is used with
skirted PCR plates, non-skirted PCR plates, standard plates, and non-configured deep well
plates.
Note: A non-configured plate is a plate that your stacker system has not been configured
to use.
Plate Type
Accessories:
Standard Plate Lane
Deep Well Plate Lane
Standard Plate
x9150
x9150 and x9152
Skirted PCR Plate
x9150
x9150 and x9153
Non-skirted PCR Plate
N/A
x9150 and x9151
Deep Well Plate
N/A
x9150
Static Plate Holder Accessories Required to Run Different Plate Types
Part Numbers
Description
X9150
Static Holder
X9151
Non-skirted PCR Plate Deep Lane Adapter
X9152
Standard Plate Deep Lane Adapter
X9153
Skirted PCR Plate (DW LANE) Adapter
Static Plate Holder Accessories
The static holder (X9150) is required to run a process using a non-configured standard plate in
the stacker standard-plate lane. To run a process using a standard-sized plate in a deep-well
plate lane, a Deep Well Lane Adapter is required to be placed on top of a static holder. The type
of adapter needed depends on the type of standard plate to be run. The adapters that are
available include: for a non-skirted PCR plate (X9151), for a standard plate (X9152), and for a
skirted PCR plate (X9153).
Note: QPix 460 systems are supplied standard with the static holder (X9150).
Summary of Contents for QPix 450
Page 1: ...5031498 G August 2022 QPix 450 or QPix 460 Colony Picking System User Guide ...
Page 6: ...QPix 450 or QPix 460 Colony Picking System User Guide 6 5031498 G ...
Page 22: ...QPix 450 or QPix 460 Colony Picking System User Guide 22 5031498 G ...
Page 56: ...QPix 450 or QPix 460 Colony Picking System User Guide 56 5031498 G ...
Page 102: ...QPix 450 or QPix 460 Colony Picking System User Guide 102 5031498 G ...
Page 158: ...QPix 450 or QPix 460 Colony Picking System User Guide 158 5031498 G ...
Page 176: ...QPix 450 or QPix 460 Colony Picking System User Guide 176 5031498 G ...
Page 218: ...QPix 450 or QPix 460 Colony Picking System User Guide 218 5031498 G ...
Page 226: ...QPix 450 or QPix 460 Colony Picking System User Guide 226 5031498 G ...
Page 250: ...QPix 450 or QPix 460 Colony Picking System User Guide 250 5031498 G ...
Page 256: ...QPix 450 or QPix 460 Colony Picking System User Guide 256 5031498 G ...