Chapter 6: Picking Processes
5031498 G
67
Running Picking Processes
After a picking routine has been configured, you can run the process on the instrument. If you
have not configured the picking routine you want to run, you must create a new picking routine
or edit an existing routine.
Note: Before running a picking process, it is important that you do the cleaning and set
up procedures in
Preparing to Run Processes on page 29
.
Some steps that are included in these procedures might not be available, depending on the
features included with the instrument and license.
To run a configured picking routine:
1. Open the Picking window. See
Opening the Picking Window on page 58
2. Select the picking routine you want to run. See
Selecting Picking Routines on page 58
If you do not need to make changes to the routine, select the Skip Steps check box before
clicking Next.
3. Review the configured settings for the routine. See
Viewing the Settings Summary on page
4. In the Settings Summary window, click Next.
5. After the Please Load Source window displays, load the source receptacles in the correct
locations on the instrument deck.
6. Close the instrument door.
7. Click Next to take a white light test image of the source receptacles. See
White Light Test Image on page 67
.
Adjusting the White Light Test Image
In the Test Image window, all detected features are viewed as potential colonies to pick. To
make sure that only colonies are detected, adjust the test image in this window using the
Exposure and Gain settings, the Detection settings, the Display settings, the Debris settings,
and the Depth settings. Each detected feature displays within a yellow ring.
To view the pixel intensity values in the image map to manually set a detection threshold or
detect areas of saturation within the image:
1. Right-click over the image map and select Intensity.
2. Move the cursor over the area of interest in the image map to display the pixel intensity
value.
Summary of Contents for QPix 450
Page 1: ...5031498 G August 2022 QPix 450 or QPix 460 Colony Picking System User Guide ...
Page 6: ...QPix 450 or QPix 460 Colony Picking System User Guide 6 5031498 G ...
Page 22: ...QPix 450 or QPix 460 Colony Picking System User Guide 22 5031498 G ...
Page 56: ...QPix 450 or QPix 460 Colony Picking System User Guide 56 5031498 G ...
Page 102: ...QPix 450 or QPix 460 Colony Picking System User Guide 102 5031498 G ...
Page 158: ...QPix 450 or QPix 460 Colony Picking System User Guide 158 5031498 G ...
Page 176: ...QPix 450 or QPix 460 Colony Picking System User Guide 176 5031498 G ...
Page 218: ...QPix 450 or QPix 460 Colony Picking System User Guide 218 5031498 G ...
Page 226: ...QPix 450 or QPix 460 Colony Picking System User Guide 226 5031498 G ...
Page 250: ...QPix 450 or QPix 460 Colony Picking System User Guide 250 5031498 G ...
Page 256: ...QPix 450 or QPix 460 Colony Picking System User Guide 256 5031498 G ...