Chapter 4: Starting and Setting Up the Instrument
5031498 G
29
Preparing to Run Processes
Molecular Devices recommends regular and thorough cleaning and maintenance of the
QPix 450 or 460 system Instrument to ensure that it continues to function correctly. See
Before running a process, do the following procedures.
1. Manually clean the instrument interior. See
Cleaning the Instrument on page 228
.
2. Install and fill the wash baths with a suitable quantity of the defined solutions.
Note: Before running a process that uses the wash baths, make sure that you verify
the locations and solutions of the wash baths and also the wash cycles that are
required for their use. Always top off your wash baths with their defined cleaning
fluids and then wipe down the surfaces of the instrument interior.
3. Start the instrument and the software.
4. Install the correct picking head for your application. See
.
5. Do a Sanitise process. See
Running Sanitise Processes on page 49
.
6. Do a UV Sanitise process. See
Running UV Sanitise Processes on page 50
7. Prepare plates, QTrays, OmniTrays, or Petri dishes based on the requirements for your
application.
8. Install the stacker cassettes, if applicable. See
Setting Up Stacker Cassettes on page 32
.
9. Select and run the desired process. For information about selecting, defining, and running
processes, see the following topics:
Picking Processes on page 57
Regional Picking Processes on page 119
Control Plate Creation Processes on page 159
Plating Processes on page 177
Replication Processes on page 189
Rearraying Processes on page 199
Gridding Processes on page 207
Setting Up Wash Baths
The Instrument contains three wash baths on the right side of the instrument bed. The wash
baths are used for the Sanitise process and the Sanitise profiles used with most other
processes. See
Sanitizing the Instrument Interior on page 49
.
Note: Before running a process that uses the wash baths, make sure that you verify the
locations and solutions of the wash baths and also the wash cycles that are required for
their use. Always top off your wash baths with their defined cleaning fluids and then wipe
down the surfaces of the instrument interior.
Fill the wash baths with fluid until the brushes are covered and configure the baths correctly.
For a typical picking process, put the following solutions in the indicated wash baths:
Wash Bath 1 (furthest from front): 70% ethanol
Wash Bath 2 (middle wash bath): Sterile water
Summary of Contents for QPix 450
Page 1: ...5031498 G August 2022 QPix 450 or QPix 460 Colony Picking System User Guide ...
Page 6: ...QPix 450 or QPix 460 Colony Picking System User Guide 6 5031498 G ...
Page 22: ...QPix 450 or QPix 460 Colony Picking System User Guide 22 5031498 G ...
Page 56: ...QPix 450 or QPix 460 Colony Picking System User Guide 56 5031498 G ...
Page 102: ...QPix 450 or QPix 460 Colony Picking System User Guide 102 5031498 G ...
Page 158: ...QPix 450 or QPix 460 Colony Picking System User Guide 158 5031498 G ...
Page 176: ...QPix 450 or QPix 460 Colony Picking System User Guide 176 5031498 G ...
Page 218: ...QPix 450 or QPix 460 Colony Picking System User Guide 218 5031498 G ...
Page 226: ...QPix 450 or QPix 460 Colony Picking System User Guide 226 5031498 G ...
Page 250: ...QPix 450 or QPix 460 Colony Picking System User Guide 250 5031498 G ...
Page 256: ...QPix 450 or QPix 460 Colony Picking System User Guide 256 5031498 G ...