Troubleshooting
SPECTRAmax 340PC Microplate Spectrophotometer Operator’s Manual
5-5
For all other error messages (codes not listed here), please contact your local
Molecular Devices representative for assistance.
Warning
Messages
Warning
messages are printed below data on the printout. A warning message
indicates a situation that requires attention but is not sufficient to stop or prevent
a reading, such as low memory, entries being out of range, or operations that
could result in loss of data. Warning messages are generally self explanatory. For
assistance regarding warning messages, contact your local Molecular Devices rep-
resentative.
Opening the
Drawer
Manually
If an error occurs while the drawer is closed and you need to remove a microplate,
press the
[Drawer]
key. If the drawer does not open, turn power to the instru-
ment off and then on again.
If the drawer still remains closed, turn the power off and, using your thumbnail,
locate the groove in the upper left side wall of the door. Open the door and use
your index finger to pull the microplate drawer out of the instrument
(do not force
the drawer)
and remove the microplate. This action will not harm the instrument,
but should only be taken if the first two options have failed to open the drawer.
If you are still unable to open the drawer, contact your local Molecular Devices
representative.
301
safe temperature limit
exceeded
A temperature of over 50°C was detected on one or more of the 4
thermistors. Temperature will be shut off and remain off until
power up reset is completed successfully.
302
low light measurement
Not enough light detected to make an accurate measurement.
310
Signal level saturation
Make sure nothing is blocking the door from shutting com-
pletely.
317
LCD time-out
LCD front panel is bad or disconnected.
400 -
499
Errors in Non-Volatile Mem-
ory
Contact Molecular Devices Technical Support.
Table 5.1: Error Codes, Messages, and Notes about the Errors
Error
Code
Error Message
Notes
Summary of Contents for SPECTRAmax 340PC
Page 4: ...iv SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Page 10: ...Chapter 1 1 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Page 30: ...Chapter 1 1 22 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Page 32: ...Chapter 2 2 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Page 36: ...Chapter 2 2 6 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Page 38: ...Chapter 3 3 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Page 44: ...Chapter 3 3 8 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Page 46: ...Chapter 4 4 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Page 52: ...Chapter 4 4 8 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Page 54: ...Chapter 5 5 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Page 58: ...Chapter 5 5 6 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Page 60: ...Appendix A A 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Page 62: ...Appendix B B 2 SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...
Page 68: ...ii SPECTRAmax 340PC Microplate Spectrophotometer Operator s Manual...