SpectraMax i3x Multi-Mode Detection Platform Installation Guide
12
5025026 C
To connect a gas supply:
1. Power off the instrument.
2. Use a flat head screwdriver to pry off the small black cap on the rear of the
instrument along the right edge to access the quick connect fitting. Save the cap for
later use.
3. Connect one end of the tubing to the quick connect fitting and connect the other end
to the gas supply.
4. Adjust the gas supply input pressure.
Note:
For the maximum permitted air supply pressure for the gas inlet and
polyurethane tubing specifications, see the user guide for the SpectraMax i3x.
To disconnect the polyurethane tubing:
1. Power off the instrument.
2. Turn off the gas supply at the source and wait a sufficient time for the pressure to
dissipate.
3. Use the gas inlet unlock tool to press the quick connect fitting and release the tubing.
Note:
The gas inlet unlock tool is not supplied with the instrument.
4. Remove the tubing from the quick connect fitting.
5. Replace the black cap over the quick connect fitting.