4.0 REPLACING THE BATTERY
To replace the ba
Ʃ
ery the top or “lid” of the logger enclosure needs to
be unscrewed (Fig 6) from the bo
Ʃ
om. There are two
fl
ats on the “lid” to
facilitate the use of a wrench. Once the ba
Ʃ
ery is exposed it may be re-
moved and replaced. Note the posi
Ɵ
ve terminal when replacing (Fig 7).
Use only a 3B5700
½
AA Lithium Sulfuryl Chloride 3.9V 800mAh High
Temperature (150°C) ba ery. Monarch Part No 5396-9934 or similar.
If the new ba
Ʃ
ery has been inserted correctly, the LED on the opposite
side of the board will blink twice within 5 seconds (Fig 8). Replace the
bo
Ʃ
om and be sure to
Ɵ
ghten it completely to ensure a good seal. The
O-ring should not create a gap between the top and the bo
Ʃ
om.
4
Figure 6-Counterclockwise Figure
7-Ba
Ʃ
ery Polarity
Figure 8-LED Blinking
FPO
For posi
Ɵ
on Only
New graphics to come.
FPO
For posi
Ɵ
on Only
New graphics to come.
Inserted
Not Tight Enough
Figure 9-Good Seal
X
LED blinking 2X
√
5.0 SPECIFICATIONS
Measurement
Temperature Range
(fixed probe):
-50° to +150°C (-58° to 302°F)
Accuracy:
+0.1°C from -30° to +150°C (+0.2°F from -22 to 302°F)
+0.2°C from -50°C to -30 (+0.4°F from -58°F to -22°F)
Temperature Range
(flexible probe):
-70° to +300°C (-94° to 572°F) (
Probe ONLY
)
Accuracy:
+0.1°C from -30° to +150°C (+0.2°F from -22° to 302°F)
+25°C from -70° to +300°C (+0.5°F from -22° to 572°F)
Environmental
Temperature Range:
-50° to +150°C (-58 to 302°F)
Humidity:
0 to 100% condensing
Pressure:
0 to
9 bar (0 to 130 psi)
Submersion:
IP68 – 1.5m (~ 5 feet) of static water indefinitely
Ba ery:
3B5700 ½ AA Lithium Sulfuryl Chloride 3.9V 800
mAh rated 150°C. Field Replaceable Monarch
P/N 5396-9934
Life: Up to 2 years typical @ 1 minute sample rate
(@20°C)
Enclosure:
Material: 316L stainless Steel.
Dimensions: 3.6cm (dia.) x 5.5cm / 1.43” (dia.) x 2.30”
(Excl. probes)
Probe Lengths: 2.5cm, 5cm, 10cm, 56cm flexible.
1”, 2”, 4”, 22” Flexible.
General
Sample Rate:
User configured, 1 sample every 2 seconds to 1 every
24 hours
No. of Samples:
30,000 stored in non-volatile memory.
Record Trigger:
Two independent triggers. Multiple trigger modes:
Instantaneous, on alarm, time and date
(start and stop).