4.0 REPLACING THE BATTERY 

To replace the ba

Ʃ

ery the top or “lid” of the logger enclosure needs to 

be unscrewed (Fig 6) from the bo

Ʃ

om.  There are two 

fl

ats on the “lid” to 

facilitate the use of a wrench. Once the ba

Ʃ

ery is exposed it may be re-

moved and replaced. Note the posi

Ɵ

ve terminal when replacing (Fig 7).  

 

Use only a 3B5700 

½ 

AA Lithium Sulfuryl Chloride 3.9V 800mAh High 

Temperature (150°C) ba ery. Monarch Part No 5396-9934 or similar.

 

If the new ba

Ʃ

ery has been inserted correctly, the LED on the opposite 

side of the board will blink twice within 5 seconds (Fig 8). Replace the 
bo

Ʃ

om and be sure to 

Ɵ

ghten it completely to ensure a good seal. The  

O-ring should not create a gap between the top and the bo

Ʃ

om. 

 Figure 6-Counterclockwise Figure 

7-Ba

Ʃ

ery Polarity 

Figure 8-LED Blinking 

FPO 

For posi

Ɵ

on Only 

New graphics to come. 

FPO 

For posi

Ɵ

on Only 

New graphics to come. 

 

Inserted 

Not Tight Enough  

Figure 9-Good Seal 

LED blinking 2X 

 

 

 

5.0 SPECIFICATIONS 

Measurement 
Temperature Range

   

(fixed probe): 

 

-50° to +150°C (-58° to 302°F) 

Accuracy: 

+0.1°C from -30° to +150°C (+0.2°F from -22 to 302°F)  

 

+0.2°C from -50°C to -30 (+0.4°F from -58°F to -22°F) 

Temperature Range  
(flexible probe):  

-70° to +300°C (-94° to 572°F) (

Probe ONLY

Accuracy: 

+0.1°C from -30° to +150°C (+0.2°F from -22° to 302°F) 

 

 

+25°C from -70° to +300°C (+0.5°F from -22° to 572°F) 

Environmental 
Temperature Range: 

-50° to +150°C (-58 to 302°F)

 

Humidity: 

0 to 100% condensing 

Pressure: 

0 to

 

9 bar  (0 to 130 psi) 

Submersion: 

IP68 – 1.5m (~ 5 feet) of static water indefinitely  

Ba ery:

 

3B5700 ½ AA Lithium Sulfuryl Chloride 3.9V 800 
mAh rated 150°C. Field Replaceable Monarch  
P/N 5396-9934

 

 

Life: Up to 2 years typical @ 1 minute sample rate 
(@20°C) 

Enclosure: 

Material: 316L stainless Steel. 

 

Dimensions: 3.6cm (dia.) x 5.5cm / 1.43” (dia.) x 2.30”  
(Excl. probes) 

 

Probe Lengths: 2.5cm, 5cm, 10cm, 56cm flexible.   
1”, 2”, 4”, 22” Flexible. 

General 
Sample Rate:  

User configured, 1 sample every 2 seconds to 1 every 
24 hours 

No. of Samples:   

30,000 stored in non-volatile memory. 

Record Trigger:   

Two independent triggers. Multiple trigger modes: 

 

 

Instantaneous, on alarm, time and date  

(start and stop). 

www.

.com

information@itm.com

1.800.561.8187

Reviews: