Connect to your data logger 

If the USB drivers installed properly you will see “Device Connected” 
displayed in the lower right hand corner of the Track-It™ Software 
window.  
 
 
 
 
 
 
 
 
You can now begin using Track-It™ Software to configure your data 
logger and/or view recorded data. 
 
Track-It™ Software will list all Track-It™ Data Loggers connected to 
your PC. To view the list and switch between loggers click on Devices 
then Device List. 

Configure and start logging data 

Select the “

Device Setup

” tab as shown. The software will 

automatically retrieve and display the data logger configuration. 

Review default factory configuration settings and make any  
necessary changes - sample rates, triggers, alarms etc. Be sure that 
the time and date is set properly. 
Click the “

Send To Device

” button under the Device Configuration 

section of the window. 

Your data logger is now operational! 

LED INDICATORS  (for units without displays) 

LCD DISPLAYS 

Some Track-It data loggers are equipped with LCD displays. The display is 
user programmable and can display real time information including time, 
date, measured values with engineering units, battery level, alarm        
conditions, etc. LED indicators on units with LCD displays are also user       
programmable. See operation manual for full description and set up    
instructions. 

Bell

 

On steady = alarms enabled. Blinks if any alarm detected 

Up arrow = Hi alarm, Down arrow = Lo alarm.  Blinks if any alarm is 
current. On steady if any alarm has been detected but not current. 

REC On steady = Triggered to record but not currently recording. 
Blinking = recording data 

Battery Condition: Indicates full, half and empty. Blinks = too low to 
function reliably. 

Icon Indications 

LED indicators are provided to notify the user of data logger status.    
Pushing and releasing the data logger button will cause the LED to flash in 
various patterns to indicate status. Pressing and holding the data logger 
button will change record mode or reset alarms (these features need to be 
configured in the software prior to use). The table below shows the push- 
button / LED functionality. 

 

STATUS 

 

   LED Sequence 
 
Low battery - one 

RED

 blink (No red if battery OK) 

 
Recording or triggered to record - 

fast 

GREEN

/

RED

 blink 

 
Idle/Not recording - one long 

GREEN

 blink 

 
Alarm occurred - one long 

RED

 blink (No red if no alarm) 

 

Not  set up or memory full - 

slow 

RED

/

GREEN

 blink, connect to PC 

 

RESULT OF BUTTON RELEASE 

 
LED pattern 
 
Release button while LED GREEN - Turn record mode on/off* 
 
Release button while LED RED - Reset alarm indication* 
 
Release button when LED goes off - No change 
 

 

*NOTE: Feature must be enabled in software 

STATUS - Press & Release Button 

 