Monark 928 G3 / 928 G3r
11
Fig: Display
1) START STOP
2) ENTER
3) (+/-) button
4) HEART button (pairing meter and chest belt)
5) PROGRAM
6) Metronome (on the back)
Base Power
(Except 928 G3r)
Default value which is used as "idle" power when no
program or test is active. The default value is 25 but
can be adjusted with (+/-) button.
Steady State HR
(Except 928 G3r)
Default value is (±)3 but can be adjusted between 1
and 10 with (+/-) button.
Steady State (SS) HR is used only in Astrand program.
SS is checked by comparing the mean value of HR
during the period 4:45 to 05:00, and the mean time
between 5:45 to 6:00. If the difference between these
two values is the same or less HR is counted in SS.