EnDaL curve/vi
PC EVALUATION SOFTWARE ENCUR
- 23 -
N°
Label
Function
1
Meas. channels
(cycl.)
Activates/deactivates the cyclic measured parameters (Interval
T/F/AN):
GPS, temperature (T), relative humidity (H), analogue channel (AN)
2
Events
(Digital)
Activates/deactivates the event triggered parameters:
Digital channels (D0, D1, D2, D3),
3
Retain sensor settings
The settings of the sensor configuration window and the parameters
inner frame (measuring range and acceleration channels) won’t be
transferred to the device, if check. Deactivation of the checkbox opens
the window for sensor configuration.
4
Events
(Acceleration)
Activates/deactivates the event triggered parameters:
Acceleration channels (X, Y, Z),
GPS for digital channels and acceleration channels.
From ENCUR 3.0, the meas. channels is only configurable using the
sensor configuration window.
5
Meas. range
Shows the measuring range of the sensor and the corresponding
maximum acceleration value. From ENCUR 3.0, the meas. range is
only configurable and transferable using the sensor configuration
window.
6
Reg. threshold
The percentage threshold of measuring range at which an acceleration
event will be recorded
7
Alarm threshold
The percentage threshold of measuring range at which an alarm event
will be signalized.
8
Frequency range
Configures the cutoff frequency for the low pass filter of 4th order in
discrete steps between 16 Hz and 1000 Hz.
9
Min. shock duration Minimal shock duration at which an acceleration event will be recorded
10
Curve duration
Duration of the recorded acceleration curves
(20 curves at 1024 ms, 10 curves at 2048 ms)
11
Interval T/H/AN
Interval for cyclic measurement (synchronous interval)
12
Serial number
Manufacturer serial number (read only)
13
Mailbox
Short note e.g. a transport name or route information (max. 64
characters)
14
New battery
Configures battery type and resets counter of operating hours
15
Time configuration
Sets the device time. A time zone information can be configured
16
Activation periods
Periods of time when the device is ready for measuring.