Power Reqs:                 115 VAC (+15%) or 230 VAC (+15%), 50/60 Hz

Power Consumption:    6 VA continuous, 50 VA intermittent

Altitude:                         6562 ft (2000m) max

Relative Humidity:         Suitable for Outdoors

Installation Category:    

II

Pollution Degree:          4, Suitable for indoor/outdoor use

Operating Temp:           -40° F to 150° F (-40° C to 65° C)

Max. Int Bin Temp:        Up to 300°F (149°C) w/use of bare SS cable;

                                     Up to 200°F (93°C) w/use of nylon or 

                                     polyethylene jacketed cables

Max. Vessel Pressure:  15 PSI

Measurement Range:   150 feet maximum (45.7 m)

Measurement Rate:      1.0 foot/second (typical) (0.3 m/s)

Accuracy:                      + 0.25% (distance or level)

Repeatability:                0.1 feet (30 mm)

Resolution:                    1/100 foot (3 mm)

Mounting:                      Freeze-Resistant “K” flange, 8” dia. 

                                     w/ 7” bolt circle

Conduit Entry:               (2) 3/4” NPT

Cable:                            1/16" nylon-jktd (270 lb/123 kg); 3/64" SS 

                                     unjktd (270 lb/123 kg); 1/16" polyethylene-jktd

                                     (270 lb/123 kg)

SMU Weight 

(Approx.)

:      42 lbs (19.1 kg)

Plumb Bob Weight:          2 lbs (0.9 kg)

Shipping Weight 

(Approx.)

:  45 lbs (20.4 kg) - [Includes SMU, plumb bob, 

                                     cable and carton]

Ship Carton Dimensions: 34 x 15 x 16 inches (864 X 381 X 406 mm) - 

(L X W X H)                   [Includes SMU, plumb bob & cable]

Enclosure/Housing:       Cast Aluminum, powder coat finish

Air Purge Connection:  1/4” NPT 

Protection:                     NEMA 4X / Enclosure Type 4X; IP66

Approvals:                     CSA

us/c

:  Ordinary Locations;

                                                    Class 

II

, Group E,F,G; Class 

III

                                     ATEX:    

II

1/2 D c

                                                    Ex tb lllC T75

o

C Db 

I

P66

                                                    (Ta -40

o

C TO +65

o

C)

                                     IECEx:   Ex tb lllC T75

o

C Db 

I

P66

                                                    (Ta -40

o

C TO +65

o

C)

                                     CE Mark

                                     CHINA RoHS 2

RS-485 Model

Input/Output:           

RS-485, half-duplex, isolated, proprietary protocol

Address Selection: 

1-16 (SiloPatrol); 1-247 (Modbus)

Wiring Distance:     

3280 ft (1000 m) @ 9600 baud, 26 AWG

Relay Output:    

5A @ 250VAC max, isolated, dry contact

Sound/Lockout:            (Vhi)9 8-265VAC/DC, 9.5K ohm input 

resistance, isolated; (Vlo) 20-55 VAC/DC,   

2.0K ohm input resistance, isolated

Analog(4-20mA) Model

4-20mA:            

Reversible to 20-4mA, self powered 500 ohm 

max loop, isolated

Relay Output:    

5A @ 250VAC max, isolated, dry contact

Sound/Lockout:            (Vhi) 98-265VAC/DC, 9.5K ohm input 

resistance, isolated; (Vlo) 20-55 VAC/DC,  

2.0K ohm input resistance, isolated

SPECIFICATIONS

Monitor Technologies LLC warrants each 

Silo

Patrol

®

SE

SMU inventory monitoring sys-

tem it manufactures to be free from defects in material and workmanship under normal

use and service for two (2) years from the date of purchase.   The purchaser must notify

Monitor of any defects within the warranty period, return the product intact, and prepay

transportation charges.  The obligation of Monitor Technologies LLC under this warranty

is limited to repair or replacement at its factory.  This warranty does not apply to any prod-

uct which is repaired or altered outside of Monitor Technologies’  factory, or which has

been  subject  to  misuse,  negligence,  accident,  incorrect  wiring  by  others,  or  improper

installation.  Monitor Technologies LLC reserves the right to change the design and/or

specifications without prior notice.

WARRANT Y

SAFET Y

       

General Safety 

CAUTION:

It is essential that all instructions in this manual be followed to

ensure proper operation of the equipment and safety of operating personnel.

The use of this symbol is used throughout this manual to highlight important

safety issues.  Please pay particular attention to these items.

        Electrical Shock Caution

SMUs are powered with HIGH VOLTAGE.  No operator serviceable parts are

inside.  All servicing is to be performed by qualified personnel.  Each SMU is pro-

vided with a "protective conductor terminal"      which shall be terminated to earth

ground potential (See Electrical Installation).  This product's design complies with

EN61010-1 installation category II and pollution degree 2.

        Hazardous Location Caution

Certain SMUs can be used in Hazardous Locations (See Specifications).

These models shall only be used in applications covered by stated ratings or

those considered non-hazardous.  Failure to comply could result in damage

to personnel and property.  The following must be maintained to assure safe

operation:

1)   Enclosure integrity

- the dimensions of the housing, side covers, 

      cover retention hardware, or through shafts shall not be altered.

2)  Maintenance

- Power to all circuits must be disconnected before con-

     ducting any investigation, setup, or maintenance when in a hazardous 

     atmosphere.

        Mechanical Hazards  

Consider the SMUs weight whenever lifting or moving the instrument.  When

being  serviced  by  authorized  personnel,  extreme  caution  shall  be  taken

whenever power is supplied to a SMU unit and the cable and associated

mechanisms are accessible.  The gear motor has sufficient torque to pinch

fingers between the cable and other mechanical parts.  Particular areas of

danger  include  the  storage  reel,  measuring  wheel,  rollers,  and  take-up

spring assembly.

Electromagnetic Compatibly (EMC)

The SMU was tested and found to comply with the standards listed below.

The SMU should not be used in residential or commercial environments.

Compliance to the EMC standards was demonstrated by means of a test

setup using the following installation methods.

1)

  SMU enclosure was connected to earth ground (protective earth).

2)

  Shielded cable was used to interconnect the RS-485 network (connec-

     tions +D and -D).  The shield drain was connected to (SHD terminal) 

     at each end.

EMC Emissions:

Meets    EN 61326-1         Electrical Equipment for Control Use, EMC

              EN 55011             Radiated and conducted emissions 

                                           (Class A - industrial)

              EN 61000-2         Harmonic Current Emissions

              EN 61000-3         Voltage Fluctuation & Flicker

Meets    FCC Part 15B      RF Devices, Unintentional Radiators

              CISPR 11             Radiated and conducted emissions 

                                           (Class A- industrial)

EMC Immunity:

Meets    EN 61326-1         Electrical Equipment for Control Use, EMC

              EN 1000-4-2        Electrostatic discharge (industrial)

              EN 1000-4-3        RF radiated EM fields (industrial)

              EN 1000-4-4        Electrical fast transients (industrial)

              EN 1000-4-5        Electrical surges (industrial)

              EN 1000-4-6        RF conducted EM energy (industrial)

              EN 1000-4-8        Power frequency magnetic fields (industrial)

              EN 1000-4-11      Source voltage deviation

B U L L E T I N

344A

GDR

JSC

AUG 04 2017

DOC-001-344A      24

Teflon

®

is a registered trademark of Dupont Chemical Co.

Bluetooth

®

is a registered trademark of Bluetooth SIG, Inc.

Android, Google Play, and the Google Play logo are trademarks of Google Inc.

Modbus

TM

is a trademark of Schneider Electric / Modicon

Monitor Technologies LLC

44W320 Keslinger Rd.  

t

Elburn, IL 60119

t

800-766-6486 (630-365-9403)   

t

E-mail: customerservice@monitortech.com   

t

www.monitortech.com

Summary of Contents for Silo Patrol SE 344A

Page 1: ...label on the SMU cover Verify the voltage output type cable type cable length flange style and location approval ordinary or haz ardous All SMUs are shipped with a standard plumb bob to ease instal lation If an alternate plumb bob is ordered it will be packed separately See Mechanical Installation section for method to change plumb bob In the event that information contained herein does not comple...

Page 2: ...res the amount of cable dispensed via its unique optical sensing system The SMU s optic system is completely sealed from the internal environment of the electronics compartment which is isolated and sealed from material ingress in the mechanical compartment to ensure long term reliable operation The descent of the bob is maintained at an optimal speed by the smart motor control system which contri...

Page 3: ...ECT ELECTR ELECTRI ELECTRIC ELECTRICA ELECTRICAL ELECTRICAL ELECTRICAL I ELECTRICAL IN ELECTRICAL INS ELECTRICAL INST ELECTRICAL INSTA ELECTRICAL INSTAL ELECTRICAL INSTALL ELECTRICAL INSTALLA ELECTRICAL INSTALLAT ELECTRICAL INSTALLATI ELECTRICAL INSTALLATIO ELECTRICAL INSTALLATION FACT ELECTRICAL INSTALLATION F ELECTRICAL INSTALLATION FA ELECTRICAL INSTALLATION FAC ELECTRICAL INSTALLATION FACTO EL...

Page 4: ...G DIAGRAM SMU RS 485 MODELS ADV WIRING DIAGRAM SMU RS 485 MODELS ADV M WIRING DIAGRAM SMU RS 485 MODELS ADV M0 WIRING DIAGRAM SMU RS 485 MODELS ADV M00 WIRING DIAGRAM SMU RS 485 MODELS ADV M006 WIRING DIAGRAM SMU RS 485 MODELS ADV M0068 7 WIRING DIAGRAM SMU RS 485 MODELS ADV M0068 7 WIRING DIAGRAM SMU RS 485 MODELS ADV M0068 7 WIRING DIAGRAM SMU RS 485 MODELS ADV M0068 7 2 WIRING DIAGRAM SMU RS 48...

Page 5: ...OU LOCKOUT LOCKOUT N L N N L1 N L1 POWER N L1 P N L1 PO N L1 POW N L1 POWE N L1 POWER DEPL N L1 POWER N L1 POWER D N L1 POWER DE N L1 POWER DEP N L1 POWER DEPLU N L1 POWER DEPLUGGABL N L1 POWER DEPLUG N L1 POWER DEPLUGG N L1 POWER DEPLUGGA N L1 POWER DEPLUGGAB N L1 POWER DEPLUGGABLE N L1 POWER DEPLUGGABLE N L1 POWER DEPLUGGABLE FROM BOT N L1 POWER DEPLUGGABLE F N L1 POWER DEPLUGGABLE FR N L1 POWER...

Page 6: ... 485 network SMU Tx orange When illuminated indicates that the SMU is transmitting data out onto the RS 485 network Status yellow When flashing indicates the general well being of the on board Bluetooth radio When flashing rapidly indicates that the Bluetooth radio is currently being pro grammed and is not available for Bluetooth connections nor is the SMU able to take a sounding Off indicates the...

Page 7: ...and 21 5mA See Page 8 for the configuration process 3 Error Reaction The 4 20mA analog SMU can be config ured to produce an error signal on the 4 20mA signal Upon an operational error the SMU will produce a current value of 2mA or 23mA The receiving device must be configured with a respective threshold to recognize the signal is outside of the standard operational range 3 21 5mA and that some form...

Page 8: ... phone or tablet from Monitor s website at http www monitortech com apps or scan the following QR Code 1 l Install the app from an e mail Consult factory l If device does not have Wi Fi connection or mobile data service other options like transfering the app from a PC to the device via a USB connection might be available Consult factory App must be connected with the SMU through Bluetooth in order...

Page 9: ... dependent on the protocol selected default address 1 RELAY SETTINGS RELAY ACTION FORCE RELAY RELAY THRESHOLD OFF CYCLE OPERATE ERROR LOW THRESHOLD HIGH THRESHOLD NOT ACT on ACT off No relay action energized whenever power is applied Relay relaxed while SMU is sounding ener gized when bob idle default CYCLE OPERATE Relay relaxed when error is detected energized when error dismissed When distance t...

Page 10: ... READ ONLY Defines physical length of cable Sets max value for relay threshold cycle limit Sounding stops upon measure reaching value Also presets the time values in Descend Assist of Zone 1 Range 1Ft 150Ft typc 1Ft 200Ft special cases SMU will periodically report distance measured thus far while still sounding EXISTING MEASUREMENT This option is only viewable when Descend Assist is ON 0 008 168H ...

Page 11: ...Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory SHOW REG MAP For Modbus protocols only Displays the Modbus register map as can be see in Installation Bulletin 344N 11 Menu Selection for ConfigureSensor App RS 485 Continued Factory Defaults are Shown in Bold Denote...

Page 12: ...es relay energized Use distance from latest measure only update on good cycles relaxed if Genesis When distance thres relay energized When distance thres relay relaxed Use distance from latest measure only update on good cycles relaxed if Genesis Relay temporarily not activated on regardless of status Relay temporarily activated off regardless of status Distance threshold range 0 150ft 0 45 7m def...

Page 13: ...tion is only viewable when Descend Assist is ON CABLE LENGTH CSL READ ONLY Defines physical length of cable Sets max value for relay threshold cycle limit Sounding stops upon measure reaching value Also presets the time values in Descend Assist of Zone 1 Range 1Ft 150Ft typc 1Ft 200Ft special cases 0 008 168H ...

Page 14: ... errors were detected they will be indicated on the HMI2 The following error conditions are analyzed and dis played if applicable Com Error Communication between HMI2 and SMU is faulty Return Error The measurement of descent was greater than the ascent If 4 100mm 1 upward travel Index Error The measurement of ascent was greater than the descent If 4 100mm Motion Error The plumb bob did not move up...

Page 15: ...on cable around rollers measuring wheel and through the wiper seal compartment and mounting flange Do not install wiper seals yet 7 Reattach plumb bob Pass cable through two compression sleeves around thimble and back through compression sleeves Crimp compression sleeves 8 Install wiper seals Install wiper seals as explained in the Wiper Seal Replacement section 9 Re establish operation Remove all...

Page 16: ... The plumb bob did not travel as far when ascending as it did when descending Determine if plumb bob is trapped in material Request another measurement A subsequent motion error confirms an entrapment A subsequent index error indicates that plumb bob has freed itself Request another measurement and errors will likely cease 2 Index Error The plumb bob did not travel as far when descending as it did...

Page 17: ... IN BRACKETS Silo Monitoring Unit SMU Digestible Plumb Bob 6 2190 Standard K Flange 6 3105 5 K Flange 6 3125 10 K Flange 6 3128 Standard Aluminum Plumb Bob 6 3103 Standard Aluminum Teflon Coated Plumb Bob 6 3123 Standard Stainless Steel Plumb Bob 6 3136 Inverted Cone Plumb Bob 6 4106 Inverted Cone Teflon Coated Plumb Bob 6 4126 Ball Float Plumb Bob 6 4108 Ball Float Teflon Coated Plumb Bob 6 4128 ...

Page 18: ...manual to highlight important safety issues Please pay particular attention to these items Electrical Shock Caution SMUs are powered with HIGH VOLTAGE No operator serviceable parts are inside All servicing is to be performed by qualified personnel Each SMU is pro vided with a protective conductor terminal which shall be terminated to earth ground potential See Electrical Installation This product ...

Reviews: