FUSE TEST
WARNING
Always perform a Fuse Test before using the
internal current shunt measurement mode.
1.
Perform a resistance measurement with the C and Ω
terminals connected
2.
If the resistance measured is greater than 0.5Ω, replace
the fuse*
* The resistance mode on the Mooshimeter is not accurate below
1Ω. The actual resistance of the fuse should be 20mΩ.
FIGURE 8 FUSE TEST
FUSE REPLACEMENT
WARNING
Only use appropriately rated replacement fuse.
1.
Open the Mooshimeter per directions above.
2.
Gently remove existing fuse from holder.
3.
Replace fuse only with appropriately rated replacement.
4.
Recommended model:
Reomax 632.300.12, available from Mooshim Engineering
Size 3AG 12A hold 600V 10kA minimum interrupt
5.
Close the Mooshimeter per directions above.