MOREHOUSE INSTRUMENT COMPANY INC. MODEL 4215 PLUS V5.0.6
Model 4215 Plus
Pg. 21 PUB.6396-02
Morehouse Instrument Company, Inc.
Tel: (717) 843–0081
To turn a Window on for a Specific Channel, use the “Enter” button to progress
through the different steps.
•
First select the channel you wish to adjust the window setting for, A or B.
•
Choose to turn it “On” or “Off” in the next step.
•
If “On” is selected, you will then be presented with the previously set widow
setting. If the setting is what is desired you may chose “Yes”, otherwise
toggle the choice to “No” and proceed to setting a new value.
•
In the next two steps you will be prompted to choose your desired unit; then
enter the Window value, in that unit.
•
Pressing the “Enter” button after inputting the value will complete the Filter
Window setup with the newly Set Value.
>>Decimal Select
The >>Decimal Select lets the user set the Maximum Number of Digits to the right of
the decimal which are displayed or printed for the selected item value for each
channel.
•
When large Load Cells are selected the number of decimal digits shown may
be limited.
•
For example, a 100 Lb Cell will allow 3 digits, while a 1000 Lb Cell will only
allow 2 digits.
•
Also, if a number is too large to fit on the display some decimal digits may be
dropped (the display will “Auto-Range”).
•
The number of digits shown for position data depends only on the number of
counts-per-inch of the Encoder and is not affected by this setting.
>>Count By Select
The >>Count By Select allows the user to choose the smallest unit of counting for
the display of the selected items value. Counting by 1, 2, 5, 10, and 20 are the
options.
Note
: The operation of the Analog Outputs and the Limits are not affected by the
settings of “Decimal Select” or “Count By Select”. Thus, if you set a Limit to be activated
at 10.001 Lbf, and have the “Smallest Count” set for 20, the Limit is activated at 10.001
Lbf., even though the display is only showing 10.000.
>>Second Line
The >>Second Line Option allows the user to choose what you would like to have
displayed on the second line.
•
You can select Limits, select the second data, select text message, or leave
the line blank. When selecting text, the text string is displayed for editing.
•
Pressing the Plus (“
+
”) and Minus (“
-
“) keys at the same time will reset the
text to an “A”.