>>Vel Filter Level
The >>Vel Filter Level sets the size of the Velocity Filter Window. The Filter Time
and Settling Time for each of the 4 levels of filtering is shown in the table below:
Level
1
2
3
4
Window Time
(seconds)
1/3
2
5
10
Settling Time
(seconds)
2/3
4
10
20
Note:
These times are based on a sample rate of 60 Hz. If sampling is slower, these
times will adjust proportionately (See >>Filter Select).
>>Max Pos Decimals
>>Max Pos Decimals mimic the “Decimal Select” setting above:
•
The setting for Position allows you to specify the number of Decimals
displayed to the right of the decimal point.
•
This setting only effects the Displayed value, if the least significant digit of the
Counts per Inch falls outside the Max Position Decimal place.
For example:
•
If the Counts per Inch is set to 1000 while Pos Decimal is set to 4, the least
significant digit effected will be 0.001 and the display will show 0.0010.
•
If the counts per inch is set to 100000 and the Position Decimal is set to 3 the
display will show 0.000 until the non-displayed decimals reach a value (in this
case greater than 0.00050) where it can be rounded up to display 0.001.
>>Max Vel Decimals
>>Max Vel Decimals mimic the “Max Pos Decimals” setting above:
•
The setting for Velocity allows you to specify the number of Decimals
displayed to the right of the decimal point.
•
This setting only effects the Displayed value, if the least significant digit of the
Counts per minute falls outside the Max Velocity Decimal place.