MOREHOUSE INSTRUMENT COMPANY INC. MODEL 4215 PLUS V5.0.6
Model 4215 Plus
Pg. 29 PUB.6396-02
Morehouse Instrument Company, Inc.
Tel: (717) 843–0081
Value Command (V)
This Command is used to read the numerical value of the desired Item and unit without
altering the front panel display.
The format is: V(item number)(unit number channel A)(unit number channel B)(repeat)
Where the
Item and unit numbers
can be found using the Question Mark Command
(@001?). And the
repeat
number must be: 0, 1, or 2.
•
Choosing repeat 1 will make the tester send you the desired value once.
•
Choosing repeat 2 will make the tester send the value at approximately 0.1
second intervals.
•
Choosing repeat 0 will turn off any values that are being sent by a repeat 2
Command.
Note:
Item numbers specific to
this command only
shown in the Question Mark Command are:
50 (Cells AB), 51(Peak AB), and 52 (Valley AB).
Example 1 (User Entered):
@001V01021
Acknowledgment 1:
@ 001 Peak A 0.0111 N
Example 2 (User Entered):
@001V5000002
Acknowledgment 2:
@ 001 Load A 120.452 lbf Load B -0.05898 lbf
@ 001 Load A 125.295 lbf Load B -0.24738 lbf
.....
@ 001 Load A 134.367 lbf Load B -0.63103 lbf
Example 3 (User Entered):
@001V00002
Acknowledgment 3:
@ 001 Load A 120.452 lbf
@ 001 Load A 125.295 lbf
.....
@ 001 Load A 134.367 lbf
Print Command (P)
The Print Command (P) will retrieve a full set of readings (like pressing the front panel
print button). These will be in the units that are currently selected on the numerical
display.
The format is: P(repeat number)
Where the
repeat number
must be: 0, 1, 2, or 3.
•
Repeat 1 will return one set of readings to the Remote Terminal.