MOREHOUSE INSTRUMENT COMPANY INC. MODEL 4215 PLUS V5.0.6
Model 4215 Plus
Pg. 30 PUB.6396-02
Morehouse Instrument Company, Inc.
Tel: (717) 843–0081
•
Repeat 2 will return a set of readings approximately every tenth second (like the
Value Command) to the Remote Terminal.
•
Repeat 0 will turn off the printing that was started with repeat 2.
•
Repeat 3 will send one set of readings to the serial printer attached to the Model
4215 Plus instead of to the Remote Terminal.
Example (User Entered):
@001P1
Acknowledgment:
@ 001 Load A 45.6250 lbf
Peak A 82.3960 lbf
Vall A -0.00083 lbf
Load B 22.5600 lbf
Peak B 23.0000 lbf
Vall B -0.00001 lbf
Pos 3.500 In
Vel 0.00000 I/M
Grs A 9.81467 lbf
Grs B 8.51474 lbf
Ch A+B -0.05408 lbf
Poly A 0.00000 lbf
Poly B 0.00000 lbf
Limits 0 0 0 1
Reset Tare, Peak, Valley, and Position Command (R)
This Command allows you to reset the Tare, Peak, Valley, and Position readings.
The format is: R(tare A)(peak A)(valley A)(tare B)(peak B)(valley B)(position select)
When taring A or B will tare the load and the Polynomial. The values for each choice are
either: a 0 (don’t reset) or a 1 (reset).
Thus you can reset any combination of these readings with a single command.
Example 1 (User Entered):
@001R1000000
Acknowledgment 1:
@ 001 Reset - Tare A
Example 2 (User Entered):
@001R0111000
Acknowledgment 2:
001 Reset - Peak A Valley A Tare B