MOREHOUSE INSTRUMENT COMPANY INC. MODEL 4215 PLUS V5.0.6
Model 4215 Plus
Pg. 34 PUB.6396-02
Morehouse Instrument Company, Inc.
Tel: (717) 843–0081
User Data Area Command (UA)
The User Data Area Command (UA) is used to set the Base Area.
The format is: UA(channel)(number)#
Example (User Entered):
@001UAA1.0025#
Acknowledgment:
@ 001 Base Area Ch A is 1.00250 sq-in
User Data Length Command (UL)
The User Data Length Command (UL) is used to set the Base Length.
The format is: UL(number)#
Example (User Entered):
@001UL2.5#
Acknowledgment:
@ 001 Base Length is 2.50000 inches
Analog Output Command (A)
This Command is used to view or set the Analog Output of the Model 4215 Plus. There
are two (2) subcommands:
•
Analog Output View (AV)
•
Analog Output Set (AS)
Analog Output View Command (AV)
The Analog Output View Command (AV) has the format AV
It will return the current settings of the Analog Output.
Example (User Entered):
@001AV
Acknowledgment:
@ 001 Analog Output is Load A in lbf
10.0 Volts indicates 1000.0 lbf
0.0 Volts indicates 0.0000 lbf