MOREHOUSE INSTRUMENT COMPANY INC. MODEL 4215 PLUS V5.0.6
Model 4215 Plus
Pg. 36 PUB.6396-02
Morehouse Instrument Company, Inc.
Tel: (717) 843–0081
Sensor Select Command (SS)
The Sensor Select Command (SS) is used to select the calibration data for a stored
Load Cell using the Load Cell Serial Number.
The format is: SS(channel)(serial number)#
Example (User Entered):
@001SSA105#
Acknowledgment:
@ 001 This is the list of cell calibration data:
Ch A = S/N 105, 4.00 lbf , 4.00000 mV/v,
5.00 V , Cal on Jan11-22, 1.4564 lbfShunt
unused S/N 110, 4.00 lbf , 4.00000 mV/v,
10.00 V , Cal on Jan11-22, 1.4440 lbfShunt
unused S/N 205, 4.00 lbf , 4.00000 mV/v,
5.00 V , Cal on Jan11-22, 1.4563 lbfShunt
Ch B = S/N 210, 4.00 lbf , 4.00000 mV/v,
10.00 V , Cal on Jan11-22, 1.4439 lbfShunt
Sensor Delete Command (SD)
The Sensor Delete Command (SD) is used to remove the calibration data for a stored
Load Cell from the list.
The format is: SD(serial number)#
Example (User Entered):
@001SD205#
Acknowledgment:
@ 001 Deleted Sensor S/N 205
Ch A = S/N 105, 4.00 lbf , 4.00000 mV/v,
5.00 V , Cal on Jan11-22, 1.4564 lbfShunt
unused S/N 110, 4.00 lbf , 4.00000 mV/v,
10.00 V , Cal on Jan11-22, 1.4440 lbfShunt
Ch B = S/N 210, 4.00 lbf , 4.00000 mV/v,
10.00 V , Cal on Jan11-22, 1.4439 lbfShunt