MOREHOUSE INSTRUMENT COMPANY INC. MODEL 4215 PLUS V5.0.6
Model 4215 Plus
Pg. 44 PUB.6396-02
Morehouse Instrument Company, Inc.
Tel: (717) 843–0081
Mass Calibration. Remove the masses from the Load Cell before entering this
command since it will perform the Shunt Calibration Measurement that is recorded for
use with the Calibration Check Command.
The format is: CMP0
Example (User Entered):
@001CMP0
Acknowledgment:
@ 001 Calibrate Command - Reading for Shunt Check...
[10 second delay]
@ 001 Calibrate Command Completed
unused S/N 105, 4.00 lbf , 4.00000 mV/v,
5.00 V , Cal on Jan11-22, 1.4564 lbfShunt
unused S/N 110, 4.00 lbf , 4.00000 mV/v,
10.00 V , Cal on Jan11-22, 1.4440 lbfShunt
unused S/N 210, 4.00 lbf , 4.00000 mV/v,
10.00 V , Cal on Jan11-22, 1.4439 lbfShunt
Ch B = S/N 205, 4.00 lbf , 4.00000 mV/v,
5.00 V , Cal on Jan11-22, 1.4563 lbfShunt
Ch A = S/N 123456, 1000.0 lbf , 4.96103 mV/v,
4.96133 mV/v,
4.96127 mV/v,
4.96128 mV/v,
10.00 V , Cal on Jan11-22, 291.08 lbfShunt
Calibrate by Shunt Command (CS)
The Calibrate by Shunt Command (CS) is used to calibrate the Load Cell using the
Shunt Calibration Factor expressed in pounds. This Command will only function if you
have already issued the Calibration Begin Command (CB). Note that the Load Cell
should not have any load on it for this command to operate correctly.
The format is: CS(shunt value in Lbf)#
The example shows a calibration using a shunt value of 320.86 Lbf.
Example (User Entered):
@001CS320.86#
Acknowledgment:
@ 001 Calibrate Shunt Command - Reading...
[10 second delay]
@ 001 Calibrate Command - Reading for Shunt Check...
[10 second delay]
@ 001 Calibrate Command Completed
unused S/N 105, 4.00 lbf , 4.00000 mV/v,
5.00 V , Cal on Jan11-22, 1.4564 lbfShunt
unused S/N 110, 4.00 lbf , 4.00000 mV/v,
10.00 V , Cal on Jan11-22, 1.4440 lbfShunt
unused S/N 210, 4.00 lbf , 4.00000 mV/v,
10.00 V , Cal on Jan11-22, 1.4439 lbfShunt