MOREHOUSE INSTRUMENT COMPANY INC. MODEL 4215 PLUS V5.0.6
Model 4215 Plus
Pg. 45 PUB.6396-02
Morehouse Instrument Company, Inc.
Tel: (717) 843–0081
Ch B = S/N 205, 4.00 lbf , 4.00000 mV/v,
5.00 V , Cal on Jan11-22, 1.4563 lbfShunt
Ch A = S/N 123456, 1000.0 lbf , 4.50037 mV/v,
10.00 V , Cal on Jan11-22, 320.86 lbfShunt
Calibrate Counts-Per-Inch Command (CI)
This command has two (2) subcommands:
•
Calibrate Counts-per-Inch View (CIV)
•
Calibrate Counts-per-Inch Set (CIS)
Calibration Counts-per-Inch View Command (CIV)
Displays the current setting of the Encoder Counts-per-inch Constant.
Example (User Entered):
@001CIV
Acknowledgment:
@ 001 Encoder is 24000 Counts per Inch
Calibration Counts-per-Inch Set Command (CIS)
Allows you to change the current setting of the Encoder Counts-per-inch Constant.
Example (User Entered):
@001CIS64000#
Acknowledgment:
@ 001 Encoder is 64000 Counts per Inch
Limits Command (L)
This Command allows you to view or set the 4 contact-closure limit switches on the
Model 4215 Plus. There are four (4) subcommands:
•
Limit View (LiV)
•
Limit Set (LiS)
•
Limit Setup Escape (LE)
•
Limit Reset (LiR)
In the above list of commands, “i” is a placeholder for the Limit number
Limit View Command (LiV)
The Limit View Command (LiV) is used to view the current settings of one of the limits.
The format is: L(limit number)V
Example (User Entered):
@001L1V
Acknowledgment:
@ 001 Lim 1 NO Enabled Load A lbf Set 0.90000 Trip>Set Latch
Off Reset 0.80000
This acknowledgment line indicates that: Limit 1 is a normally open switch (NO) rather
than a normally closed switch (NC). The limit is Enabled rather than Disabled. The