Display Count By Command (DC): is used to set the smallest number that the
display will count by for the Load, Peak, and Valley measurements.
•
Display Second Line Command (D2): is used to set the second line of the display
panel to be blank or to show the Limit status.
•
Display Second Line Text (DT): is used to enter the text that would be displayed
if text is set to be displayed on the second line.
Each of these display options are described below.
Display View Command (DV)
The Display View Command (DV) is used to return the current values of all the display
options.
The format is: DV
Example (User Entered):
@001DV
Acknowledgment:
@ 001 Filter is Type II Level 1
Filter Window A is On
Filter Window A = 1.0000 lbf
Filter Window B is Off
Channel A shows 5 decimal digits
Channel B shows 5 decimal digits
Channel A counts by 1
Channel B counts by 1
Second line shows display
Line 2 Message - Stress Test Ready
Velocity Filter Level is 1
Position shows 5 decimal digits
Velocity shows 5 decimal digits
Display Filter Command (DF)
The Display Filter Command (DF) sets the Filter Type and Level.
The format is: DF(type)(level)
Where
type
is either: 1 for Type I Filters or 2 for Type II Filters.
Level
must be 1, 2, 3, or
4 depending on the level of filtering desired. (Option 4 provides the most filtering and the
slowest rise time).
Example (User Entered):
@001DF13
Acknowledgment:
@ 001 Filter is Type I Level 3
Display Decimal Command (DD)
The Display Decimal Command (DD) sets the maximum number of decimal digits that
are shown for Load, Peak, and Valley.
The format is: DD(channel)(number)
Where
channel
is either A or B and the
number
must be between zero and 5.