Sensor Delete
The Model 4215-HS can store calibration data for 4 different load cells / sensors (25 slots optional). To delete
a load cell from this list choose “>Sensor Select” from the setup mode main menu. Press the left/right or plus/
minus keys until the serial number of the cell you wish to delete is displayed. (You may need to shut off auto-id
if it is on, see the System Options menu for details). Then press both the plus and minus keys at the same time.
The Model 4215-HS will ask “Delete Sensor?” to be sure that you intend to delete this calibration data. Press
-
ing Enter will delete the cell data from the list. Pressing Escape will return to the load cell select without chang
-
ing the list.
>Calibration
This menu is used to access all methods of calibration that are supported by the Model 4215-HS. Selecting cal
-
ibration from the setup mode main menu will bring up a submenu. This submenu is used exactly like the main
menu. Press left/right or plus/minus to change the menu item. Press escape to return to the setup mode main
menu. Press enter to select the displayed menu item.
If you are re-calibrating a load cell, the Model 4215-HS will recognize it using either the autoid or a manually
entered serial number. Any time you overwrite existing load cell calibration data you will first be warned with
the “Overwrite Data?” message. Pressing enter will overwrite; pressing escape will cancel the calibration. An
-
other method of overwriting existing data is to delete the sensor first and then calibrate.
If a load cell with a new serial number (or new auto-id) is calibrated the data is added as a new entry in the
load cell list.
>>Cal Check
Selecting this item will run a calibration check on the load cell / sensor attached to Channel A or B. The Model
4215-HS will measure the shunt value of the load cell and compare it to the shunt value that was recorded
when the load cell was last calibrated. The rated load and last calibration date for the load cell are displayed
followed by the currently measured shunt value and the shunt value that was recorded when the load cell was
calibrated.
Note that the Model 4215-HS has a switch on the back panel that allows you to select either a 30K shunt or a
60K shunt. Be sure that this switch is set correctly for the load cell that you are using.
Since this shunt check electronically applies approximately half the rated load, it is important that the load cell
/ sensor is not loaded when this check is run. If you see the “Cell Overloaded” message, remove any load on
the cell and try the check again.