4215-HS Indicator
Manual
(PM-4101)
Notes:
The Set and Reset conditions for each limit are evaluated independently. If both conditions are met then Reset
will take priority over Set.
Limit status may show an asterisk if the limit is enabled but there is an error in the signal source. For example,
if the limit source is load but there are no load cells currently calibrated, or the load cell has a rating of 0.0 Lb,
or a millivolt per volt constant of zero. Another example would be if the limit source is position and zero has
been entered for the encoder counts-per-inch constant. Limits that are in this error condition will be forced to
the off position as if they were disabled.
>System Options
This menu contains some system wide parameters that are seldom changed. These are listed in a submenu
that works exactly like the setup mode main menu.
>>Printer Baud
The front panel print button and the RS232 print commands will use this baud rate for communication with
your printer. This setting has no effect on units that are not equipped with the printing option. 57.6 K baud is
commonly used on older serial printers 230.4 K baud is used on more recent serial printers.
>>Auto Identify / TEDS Capability
This allows you to turn automatic load cell identification on or off. Usually you can leave auto-id on even if
you never use automatic load cell identification. In order to use this function you must either have an au
-
to-id equipped load cell or use an in-line auto-id module attached to your load cell cable (TEDS). Meters can
be equiped with TEDS Template 40 and 41 read & write capability. Template 40 and 41 are avaialble in the
“4215-HS-Enahanced Model”.
>>Auto Zero Ch A/B
This menu lets you enable or disable auto zeroing for load measurements on Channel A or B. When enabled
this function will tare the load to zero any time that it has remained within 10 counts of zero on the display for
10 seconds..
>>Com Address
This item allows you to set the address of this unit used for the RS232 ASCII command set. Valid values are 1
through 254. If the RS232 multi-drop option is installed you must be sure that all the attached units have differ
-
ent addresses to prevent data collisions.