Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 27
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
Analog Output Command (A)
This command is used to view or set the analog output of the Model 4215-HS. There are two subcommands
Analog View (AV) and Analog Set (AS).
Analog Output View Command (AV)
The Analog View (AV) command has the format AV. It will return the current settings of the analog output.
Example:
@123AV
Acknowledge:
@123 Analog Output A is Load A in Lb
10.0 Volts indicates 500 Lb
0.0 Volts indicates 0.00 Lb
@123 Analog Output B is Load B in Lb
10.0 Volts indicates 1000 Lb
0.0 Volts indicates 0.00 Lb
Analog Output Set Command (AS)
The Analog Set (AS) command is used to change the settings of the analog output. The format is AS(channel)
(item number)(unit number)(full scale)#(zero scale)#, where the channel is the analog out put channel. The
item and unit numbers are used to determine the signal source for the analog output. The full scale number
indicates what signal level is assigned to full scale analog output of 10.0 volts, and the zero scale number indi
-
cates the signal level assigned to zero volts analog output.
Example:
@123ASA0000500.0#0#
Acknowledge:
@123 Analog Output A is Load A in Lb
10.0 Volts indicates 500 Lb
0.0 Volts indicates 0.00 Lb
@123 Analog Output B is Load B in Lb
10.0 Volts indicates 1000 Lb
0.0 Volts indicates 0.00 Lb
Sensor Command (S)
This command is used to view or select calibration data from the list of calibrated load cells maintained by the
Model 4215-HS. There are 4 subcommands, Sensor View (SV), Sensor List (SL), Sensor Select (SS), and Sensor
Delete (SD).