The final average value will remain on the display until the input is again activated or the unit is powered
down.
If you do not require isolation for your application, you can use the +12 volt source and an available GND signal
located on the Digital I/O connector (pin 26 and pin 24) to create a circuit that only needs contact open/closure
to activate and deactivate the digital input. See the application diagram below.
Application Diagram – Contact OPEN/CLOSE activation of Force Averaging