You are now ready to turn on the power. Each time power is applied to the Model 4215-HS you will see a se
-
quence of 6 messages appear on the front display. For example,
Each of these messages will be displayed for about 3 seconds. After that the front display will start showing the
current readings from the load cell and sensor.
To change what item is being displayed press the leftmost button. This will cycle through the list of items (Load
A, Peak A, Valley A, Load B, Peak B, Valley B, Position, Velocity, Limits, optionally: Avg A, Avg B).
The rightmost button will change the units that are being used. Repeated presses of this button will cycle
through the list of units available for the currently displayed item. For example, with Load A showing you
would see Lb, kg, N, PSI, MPa, Klb, kN, t, mV/V, g and then back to Lb.
#
Message
Meaning
1
ESL .X.Y.Z
This is the version number.
2
Serial # 12345
Option # 114236
This is the serial number.
This is the option number.
3
Last ID A 90437
Last ID B 12030
The serial number of the load cell on Ch A.
The serial number of the sensor on Ch B
4
Ch A Max 4.00Lb Ch B
Max 10.00In
Maximum rated load of the sensor on channel A.
Maximum rating of the sensor on channel B.
5
Ch A Cal 4.00000 mVv
Ch B Cal 4.00000 mVv
The mV/V constant used with the load cell on channel A.
The mV/V constant used with the sensor on channel B.
Note: If either of these sensors was calibrated by 5 –Point Mass, then 4 mV/V constants will be displayed here.
6
Ch A Exc 5.00 Volt
Ch B Exc 2.50 Volt
Excitation voltage used with the load cell on channel A.
Excitation voltage used with the sensor on channel B.