Universal Calibrating Machine
(UCM) Manual
(PM-5201)
Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 19
Rev. 11/2021
Next, as shown in Figure 12, place a hardened alloy steel loading ball (B) of the proper size and load rating in
the ball seat of the reference standard (C) and lower the upper yoke platen (A) until the ball seat of the platen
rests on the ball.
Figure 11:
A) Lower Boss of Reference Standard
B) Ring Alignment Bushing
C) Jack Ram
Figure 12:
A) Upper Yoke Platen
B) Hardened Steel Alloy Load Ball
C) Jack Ram Upper Boss of Reference Standard
If the reference standard is not provided with a ball seat, a suitable adapter should be used. Do not confuse a
lathe center, often found on instruments, with properly machined ball seats.
Note: Do not interchange hardened alloy steel loading balls, retainer clips with hardened alloy steel loading
balls between different reference standards. They should be matched with a reference standard and used only
with that standard.
In cases where the reference standard does not have a ball seat, proper axial alignment is difficult to achieve
when the proper accessories and adapters are not used. To more easily maintain proper axial alignment, it is
recommended that a reference standard with a ball seat be used whenever possible.