Universal Calibrating Machine
(UCM) Manual
(PM-5201)
Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 22
Rev. 11/2021
Figure 13: Morehouse Quick Change Tension Members
Quick-Change Tension Member
A single Tension Member Assembly is installed into the
UCM and multiple tension adapters, both male and
female, are interchanged to attached the desired size
instrument.
A single pair of Tension Members can re
-
main fixed in the machine and by swapping
out both male and female tension adapt
-
ers three different load cell sizes can be
calibrated in the UCM by user.
One end of the tension adapter always
has the same size thread on it so that any
adapter chosen will fit the master size Ten
-
sion Member Assembly.