8
ENGLISH
Table 3
MANUFACTURER'S GUIDE AND DECLARATION - ELECTROMAGNETIC IMMUNITY
The CN210X Scooter is intended for use in the following electromagnetic conditions.
The client or user must ensure that it is used in these conditions.
IMMUNITY TESTS
TEST LEVEL IEC 60601
CONFORMITY LEVEL
ELECTROMAGNETIC
ENVIRONMENT – GUIDE
conducted RF IEC
61000-4-6
3Vrms From 150
kHz to 80 MHz
1 Vrms
Portable and mobile
RF communications
devices should not be
used near any part of
the CN210X Scooter,
including the cables,
unless the recommended
separation distances
are observed, calculated
by the equation
applicable to the
transmitter frequency.
Recommended distances:
d=[3,5/V1]√P
from 0 Hz to 80 MHz
d=[3,5/E1]√P
from 80 MHz to
800MHz,
d=[7/E1]√P
from 800 MHz
to 2.5 GHz,
where "P" indicates the
maximum rated output
of the transmitter in
Watt (W) according
to the transmitter
manufacturer and "d"
is the recommended
separation distance
in metres (m). The
intensity of the field of
fixed RF transmitters,
as determined by an
electromagnetic survey
of the site, a could be
lower than the level
of conformity of each
frequency range. b There
may be interference
near devices with the
following symbol:
radiated RF
IEC 61000-4-3
20 V/m From 80
MHz to 2.5 GHz
1 V/m
NOTE 1: at 80 MHz and 800 MHz, apply the separation distance for the higher frequency band.
NOTE 2: these guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
a The intensity of the field for fixed transmitters, such as base stations for radio-telephones (cellulars and
cordless) and terrestrial mobile radios, amateur radio equipment, AM and FM radio transmitters and TV
transmitters may not be known precisely in theory. To establish an electromagnetic environment caused by
fixed RF transmitters, one must run an on-site electromagnetic inspection. If the field intensity measured
in the place where Nefti is used exceeds the applicable RF level of conformity indicated above, normal
functioning of the CN210X Scooter must be monitored. If abnormalities are observed, supplementary
measures may be necessary, such as re-orientation or repositioning of the CN210X Scooter.
b The field intensity over the frequency range from 150 kHz to 80 MHz should be less than 3 V/m.