▒▓
Monitor Series
Vital signs Monitor instruction Manual
- 6 -
as an adjunct in patient assessment. It must be used in conjunction with clinical signs and
symptoms.
!
Prolonged use or the patient’s condition may require changing the sensor site periodically.
The site must be checked at least every four hours to ensure adequate adhesion, circulation,
skin integrity and correct optical alignment. If the circulatory condition or skin integrity is
compromised, the sensor should be applied to a different site.
!
Use only the battery, power cable and accessories appointed by manufacturer, as other
accessories may cause improper performance or dangers.
!
Connect the monitor to a three-wire, grounded, hospital-grade receptacle if necessary..
!
By replacing the fuse, please use the safety fuse of the same type and rated fuse.
!
When connecting the monitor to any instrument, verify proper operation before clinical use.
Refer to the other device’s manual for full instructions. Accessory equipment connected to the
monitors data interface must be certified according to IEC Standard 60601-1 for electro
medical equipment. All combinations of equipment must be in compliance with IEC Standard
61601-1-1 systems requirements. To avoid potentially hazardous leakage currents, always
check the summation of leakage currents when several item of equipment are interconnected.
!
Before use the equipment, inspect whether all the cables are in good condition, the damaged
cables and connectors must be replaced. Operator should examine whether the system is in
correct working state and operating condition.
!
As with all medical equipment, carefully route patient cabling to reduce the possibility of
patient entanglement or strangulation.
!
Do not use a tape to secure the sensor to the site; this can restrict blood flow and cause
inaccurate readings. Use of additional tape can cause skin damage or damage the sensor.
!
Dispose of the device and its accessories according to applicable local regulations.
!
To avoid an electrical hazard, never immerse the unit in any liquid or attempt to clean it with
liquid cleaning agents. Always disconnect monitor from AC Main Power before performing
cleaning of maintenance.
!
If the monitor becomes accidentally wet during use, discontinue operation of the monitor until
all affected components have been cleaned and permitted to dry completely. Contact our local
representative if additional information is required.
!
The signal output part can be only connected to the computer complying with the
requirements of IEC60950.
Cautions!
!
Do not sterilize by irradiation, steam, autoclave or ethylene oxide.
!
Operation of the monitor may be affected by the use of an electrosurgical unit (ESU).
!
The system may not conform to all performance specifications if stored or used outside the
environmental specification identified in specification.
!
Alarm must be set up according to different situation of individual patient. Make sure that
audio alarm can be activated when alarm occurs.
!
Do not only depend on the alarm system, the doctor and nurse will not draw attention when
an alarm turn down or turn off.
!
Single-use accessories should never be reused.
!
If the accuracy of any measurement does not seem reasonable, first check the patient’s vital