ELECTROMAGNETIC COMPATABILITY (EMC)
14
Table X-2 Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
The equipment is intended for use in the electromagnetic environment specified in the table below. The customer or
the user of the equipment should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment:
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
+/- 2 kV for
power supply lines
+/- 1 kV for
input/output lines
+/- 2 kV for
power supply lines
+/- 1 kV for
input/output lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
+/- 1 kV differential
mode
+/- 2 kV common mode
+/- 1 kV differential
mode
+/- 2 kV common
mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% UT (>95% dip in
UT) for 0.5 cycles
<40% UT (>60% dip in
UT) for 5 cycles
<70% UT (>30% dip in
UT) for 25 cycles
<5% UT (>95% dip in
UT) for 5 sec
<5% UT (>95% dip in
UT) for 0.5 cycles
<40% UT (>60% dip in
UT) for 5 cycles
<70% UT (>30% dip in
UT) for 25 cycles
<5% UT (>95% dip in
UT) for 5 sec
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
equipment requires continued operation
during power mains interruptions, it is
recommended that the equipment be
powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE 1: UT is the AC Mains voltage prior to application of the test level.
NOTE 2: Tests verified with shielded input/output cables only.
Summary of Contents for Quinton CR60
Page 6: ...TABLE OF CONTENTS iv...
Page 8: ...TABLE OF FIGURES vi...
Page 12: ...NOTICES 4...
Page 20: ...GENERAL CARE 12...
Page 44: ...MAINTENANCE TROUBLESHOOTING 36...
Page 50: ...SPECIFICATIONS 42...
Page 56: ...RECEIVING AND INSTALLATION 48...